2
Environment
-
Avoid installing the indicator in areas of direct sunlight or high humidity
-
Avoid sudden temperature change – if this is unavoidable allow equipment to ‘soak’ at a constant
temperature for at least three hours before use
-
Ensure that steady, clean AC power is available to the unit
Remember that the installer is ultimately responsible to assure that a particular
installation will be and remain safe and operable under the specific conditions
encountered.
CONNECTIONS
For the TI-500 Plus model, the rear panel contains all connectors necessary to make the appropriate
connections to the weigh platform, printer, remote display and power supply.
For the TI-500-SS Plus model, the rear cover must first be removed to make the appropriate connections
to the weigh platform, printer, remote display and power supply. To remove the rear cover, simply remove
the screws that secure it to the enclosure and set aside.
Caution!
Disconnect power source from indicator prior to removing rear cover.
CONNECTING THE WEIGH PLATFORM
The TI-500 Plus ships with a 15 ft shielded load cell cable for connection to weigh platform’s load cell(s) or
junction box.
1.
Plug the cable’s 14-pin parallel interface connector into the load cell port on the rear
panel of the indicator.
2.
Wire the bare wires and shield to the weigh platform’s load cell(s) or junction box using
the color codes shown below.
4-wire cable
Optional 6-wire cable
Color
Wire Name
Color
Wire Name
Red
+ Excitation
Red
+Excitation
Black
- Excitation
Black
-Excitation
Green
+ Signal
Green
+Signal
White
- Signal
Yellow
- Signal
Orange
+ Sense
Brown
- Sense