Blank analysis:
Ensure that the 10ml vial is clean.
Fill the vial with 10ml of the water sample.
Fit the cap and tighten.
Place the vial in the sample chamber, making sure that the arrow marks on the instrument
and vial are aligned.
Press the zero key
Wait until the following is displayed
Sample preparation:
Add
one Nitrite LR tablet
straight from the foil to the water sample.
Crush the tablet using a clean stirring rod, until no large pieces are visible.
Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
Analysis :
Place the vial in the sample chamber, making sure that the marks on the instrument and vial
are aligned.
Press the Test key
Wait for a reaction period of
10 minutes
After the reaction period, the measurement starts and the result is displayed in mg/l Nitrite.
LOD/Tolerance
The Lower LOD is 0.01 mg/l (10ppb), upper LOD is 0.5 mg/l (500ppb).
Tolerance: ± 0.005mg/l.
Zero accepted
Prepare Test
Press TEST