12.1Troubleshooting
LCD DISPLAY
CAUSE
CORRECTIVE ACTION
Electrode dried out
Soak the pH electrode in 3M KCL
solution for 15 minutes
Soak the conductivity electrode in tap
water for a few minutes
Measured value is out of range
Check the electrode whether clogged,
dirty or broken
Incorrect calibration solutions
Using the fresh calibration solutions for
calibration
Setting value does not match
calibration solution
Reset the calibration value
pH electrode is out of service
life
Replace the pH electrode
Keypad is not working properly
Replace the batteries
DO electrolyte solution is
depleted
Refill the electrolyte solution
Zero oxygen solution is
contaminated
Replace the calibration solution