25
5.4.6 Press
▲ or ▼ key to increase or decrease displayed value according to the value of the
custom buffer, and press ENTER key to confirm
. “Calibration” indicator begins flashing.
5.4.7 Wait for the measured value to stabilis
e, the display shows “END”. Once calibration is
complete, the meter returns to measurement mode automatically.
● During the calibration process, to exit the calibration mode, press the ON/OFF key; the
meter will return to measurement mode immediately.
● The electrode slope indicator shows average slope of the pH electrode after calibration.
When the electrode or calibration result does not meet measurement requirements, the
indicator will disappear on the display.
5.5 pH Calibration Report
The meter allows checking of the calibration data of the pH electrode for diagnostic purposes.
5.5.1. Press the mR key in the pH measurement mode, the display
shows “LOC/P-01”.
5.5.2. Press
▲ or ▼ key until the display shows “ELE/P-02” (Electrode Diagnosis).
5.5.3. Press ENTER key to confirm, the meter shows the last calibration date (Format: mm-dd).
5.5.4. Press
▼ key, the meter shows “OFS” (zero-point offset potential).
5.5.5. Press
▼ key again, the meter shows the slope of each pH buffer group.
5.5.6. Press the ON/OFF key to return to measurement mode.
If the meter is not calibrated or calibration is not successful, the display will show "----" only.
MR
MR
mV
MR
%
….......Calibration Points
….......Slope
Summary of Contents for HydroCheck HC1000
Page 1: ...HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual...
Page 45: ...45 NOTES...
Page 50: ...Trace2o Ltd Rev1 0...
Page 58: ...Trace2o Ltd Rev1 0...
Page 62: ...Trace2o Ltd Rev1 0...