31
● To exit calibration, press ON/OFF key; the meter will return to measurement mode
immediately.
● Performing the conductivity calibration will simultaneously calibrate the corresponding
TDS, salinity and resistivity value.
7.5 Conductivity Calibration Report
The meter allows checking of the calibration factor of the conductivity electrode for diagnostic
purposes.
1. Press the MR key in the conductivity measurement mode, the display
shows “LOC/P-01”.
2. Press
▲ or ▼ key until the meter shows “ELE/P-02” (Electrode Diagnosis).
3. Press ENTER key to confirm; the meter shows the last calibration date (Format: mm-dd).
4. Press
▼ key, the meter shows calibration point and calibration factor (e.g., K=0.998).
5. After the browsing, press MEAS key to exit the current mode.
EXIT THE CALIBRATION:
To exit calibration, press the ON/OFF key at any point; the meter will return to measurement
mode immediately.
MR
MR
MR
……... Calibration Point
...…... Calibration Factor