![Trace2O HydroCheck HC1000 Instruction Manual Download Page 29](http://html1.mh-extra.com/html/trace2o/hydrocheck-hc1000/hydrocheck-hc1000_instruction-manual_1146893029.webp)
29
7.2.5 Press ENTER key to confirm,
“Calibration” indicator begins flashing.
7.2.6 Wait for the measured value to stabilis
e, the display shows “END”. Once calibration is
complete, the meter returns to measurement mode automatically.
7.3 Multi-Point Calibration:
7.3.1 Make sure that you have selected 2 to 5 points calibration in setup menu.
7.3.2 Repeat steps 1.2 to 1.5 above; when the first calibration point is completed, the display
will
show “CAL2”. The meter prompts you to continue with the second calibration point.
7.3.3 Rinse the conductivity electrode with distilled water. Immerse the electrode in the next
calibration solution, the meter automatically shows the currently detected calibration
standard (e.g., 12.88mS/cm).
7.3.4 Press ENTER key to confirm,
“Calibration” indicator begins flashing.
7.3.5 Wait for the measured value to stabilis
e, the display shows “CAL3”. The meter prompts
you to continue with third point calibration.
7.3.6 Repeat steps 2.3 to 2.4
above until the display shows “END”. Once calibration is
complete, the meter returns to measurement mode automatically.