Trace2o Ltd
Rev1.0
Analysis (Total Chlorine):
Place the vial in the sample chamber, making sure that the marks on the instrument and vial
are aligned.
Press the Test key
Wait for a reaction period of
2 minutes
After the reaction period, the measurement starts and the result is displayed in mg/l total
Chlorine
.
Sample preparation (Differentiated Determination):
Add
one DPD No.1 tablet
straight from the foil to the water sample.
Crush the tablet using a clean stirring rod, until no large pieces are visible.
Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
Analysis (Differentiated Determination):
Place the vial in the sample chamber, making sure that the marks on the instrument and vial
are aligned.
Press the Test key
Remove the vial from the sample chamber.
Add
one DPD No.3 tablet
straight from the foil to the water sample.
Crush the tablet using a clean stirring rod, until no large pieces are visible.
Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
Place the vial in the sample chamber, making sure that the marks on the instrument and vial
are aligned.
Press the Test key
Wait for a reaction period of
2 minutes
After the reaction period, the measurement starts and the result is displayed in:
mg/l free Chlorine
mg/l combined Chlorine
mg/l total Chlorine
LOD/Tolerance
The Lower LOD is 0.01 mg/l (10ppb), upper LOD is 6 mg/l (6000ppb).
Tolerance: ± 0.005mg/l.