36
10.1 Temperature Calibration
When automatic temperature compensation is enabled, the meter will need to be calibrated if
the temperature reading displayed differs from that of an accurate thermometer.
10.1.1. Press and hold the MODE key for 3 seconds to enter temperature calibration mode, the
display shows current temperature reading.
10.1.2. Press
▲ or ▼ key to set the temperature value.
10.1.3. Press ENTER key to confirm. Calibration is completed.