
Page 41
Specific Ion, pH and mV Troubleshooting, continued…
Unstable readings.
1.
Reference electrolyte
chamber needs to be
refilled.
2.
Reference junction blocked.
3.
Ion sensor surface or pH
glass bulb not clean.
4.
Bubble in tip of Ion sensor
or pH glass bulb.
5.
Faulty connection to meter.
6.
Reference junction not
immersed.
7.
KCl crystals around
reference junction, inside
the electrolyte chamber.
Refill electrolyte. See Ion
sensor specifications for correct
electrolyte. Use saturated KCl
for pH sensor.
Clean reference junction, as per
instructions supplied with the
sensor.
Clean Ion sensor or pH glass
bulb as per instructions
supplied with the sensor.
Flick the sensor to remove
bubble.
Check connectors. Replace if
necessary.
Ensure that the reference
junction is fully immersed.
Rinse electrolyte chamber with
warm distilled water until
dissolved. Replace electrolyte.
Inaccurate readings,
even when
calibration is
successful.
Reference junction blocked.
Clean reference junction, as per
instructions supplied with the
sensor.
Displays constant
Specific Ion
readings or pH7.00
for all solutions.
Electrical short in connector.
1.
Check connector. Replace if
necessary.
2.
Replace sensor.
Displays constant
Specific Ion
readings or pH4 to
6 for all solutions.
Ion sensor surface, pH glass
bulb cracked, or internal
damage.
Replace sensor.
19.3 Temperature Troubleshooting
Symptom
Possible Causes
Remedy
Displays “
OVR
o
c
”
when Temperature
sensor is plugged
in.
1.
Faulty temperature sensor.
2.
Faulty instrument.
Return sensor to factory for
repair or replacement.
Return to factory for repair.
Temperature
inaccurate and
cannot be
calibrated.
1.
Faulty connector.
2.
Faulty Temperature sensor.
Check the connector and
replace if necessary.
Return sensor to factory for
repair or replacement.