Page 25
9.2 Failed Calibration
If calibration has failed, the GLP function will reset the date and time to zero. The
WP-90
still shows the results of the last successful calibration. These calibration
values are still used if further measurements are taken without re-calibrating. For
example…
Ion Offs 10.0
@ 00/00/00 00:00
Ion Slope 99.0%
@ 31/2/98 11:10
pH Asy 0.10pH
@ 31/12/98 11:20
pH Slope 99.0%
@ 31/12/98 11:30
Temp Offset 1.0
O
c
@ 31/12/97 11:40
9.3 Printing GLP Information to the RS232 Port
The GLP information stored in the instrument’s memory can be sent to a printer or
PC via the RS232 port. This function is available only when the optional RS232
port is fitted.
Switch the meter on.
Ensure that the
WP-90
RS232 cable is connected to the instrument and to the
printer or PC.
Select the GLP menu. (
→
→
F4:Setup
→
→
F3:GLP
)
Select
F3:Print
.
The GLP information is sent to the RS232 port in formatted ASCII text.
For example…
WP90 V1.0 R1234 @ 31/12/98 12:00
Ion Offset= 10.0mV @ 31/12/98 11:00
Ion Slope= 99.0% @ 31/12/98 11:10
pH Asy= 0.10pH @ 31/12/98 11:20
pH Slope= 99.0% @ 31/12/98 11:30
Temperature Offset= 1.0oC @ 31/12/98 11:40
ENDS