Page 16
8.
Rinse the Sensors in distilled water and blot dry.
Secondary Calibration
9.
Place the Ion Selective, Reference and Temperature sensors into the
Secondary Standard.
Ensure that any Ionic Strength Adjusting Buffer (ISAB) that may be required
has been added (see the Ion Selective Electrode’s handbook).
A magnetic stirrer providing constant, gentle stirring generally improves
stability.
10.
Select Secondary Calibration (
→
→
F1:Cal
→
→
F3:Ion Sec
).
The following message should be displayed. The current reading is displayed
on the top left. A “
∗
” in the reading indicates that the
WP-90
is currently not
calibrated.
0*123ppM 25.0
O
c
F1
:
Cal at0.100ppM
When the reading has stabilised, press
to perform a Secondary
Calibration.
A “
∗
” in the display will be replaced by a decimal point after a successful 2-
point Slope calibration.
11.
The WP-90 is now calibrated ready for Specific Ion measurements.