Page 35
13. Battery Saver Function
The
WP-90
is equipped with a battery saver function. If no button has been
pressed for five minutes, the unit beeps and flashes the display for 20 seconds, and
then shuts off. This function can be switched off for continuous use.
To enable or disable the battery saver function…
1.
Switch the meter on.
2.
Select Battery Saver Set-up (
→
→
F4:Setup
→
→
F1:Batt
)
The battery saver menu is now displayed…
Batt Saver
F1
:
OFF
100%
→
→
F2
:
ON
The arrow indicates the current selection.
The bar graph and percentage indicate the approximate level of charge in the
battery.
3.
Press
to disable the battery saver function for continuous use.
Press
to enable the battery saver function. The meter will switch itself
off if no key has been pressed for five minutes.
Press
to quit the battery saver menu and retain the current setting.
4.
For troubleshooting purposes, the battery volts can also be displayed in the
battery saver menu. Press
to display battery volts.
14. Recharging the Battery
The
symbol flashes when the battery drops below 5.60 volts. The battery
should be recharged at this point. If the battery is not recharged, the
WP-90
will
switch itself off when the battery drops below 5.00 volts.
To recharge the battery…
1.
Plug the battery charger, solar panel, or car cigarette lighter adaptor into the
Charger/RS232
socket.
DO NOT
plug into the
Temperature
socket, as
this will damage the
WP-90
.
2.
Charge for approximately 8 hours for full capacity. The
WP-90
has special
circuitry to prevent overcharging, so the charger can be used continuously.
3.
To ensure optimum battery life and capacity, the
WP-90
should only be
charged once the
symbol starts to flash.