7.4 ASF with Automatic Datalogging
Using ASF with Automatic Datalogging allows continuous logging until all readings are stable.
This is useful in titrations etc, where logging is no longer required once an end point is reached.
See section 11.2 for details on programming the Automatic Datalogging function.
1. Press
then
in normal measurement mode. All active parameters are highlighted and the
symbol appears in the display, along with the message “Auto Log”. For example…
99.8
%S
25.0
O
C
2760.
uS
31/12/01
Log#1
Auto
Log
12:00:00
As each parameter becomes stable, the highlighting is removed. If any parameter subsequently
becomes unstable, the highlighting is applied again.
The highlighting is provided to give the operator a visual indication of exactly which
parameters are not yet stable in case user intervention is required.
2. When ALL parameters have become stable, all highlighting will be removed and the will
change to a . All readings are now frozen and automatic datalogging has stopped. For
example…
99.8
%S
25.0
O
C
2760.
uS
31/12/01
12:00:00
3.
To unfreeze the display in order to take the next reading, press
and
again to re-
start the ASF automatic datalogging process.
4. Press
a second time, while the symbol is being displayed, to turn ASF off.
8. Display Backlight
The
smart
CHEM
-D
is fitted with an Electro-Luminescence (“EL”) backlight. This can be set up
according to your preferences as follows…
1. Switch the meter on.
2. Select the Backlight menu (
→
→
F4:Setup
→
→
F3:Backlight
).
3. The Backlight menu is now displayed…
BACKLIGHT
F1:On F2:Off
>F3:Energy Saver
The arrow indicates the current selection.
Press
to set the Backlight to be on continuously.
Press
to set the Backlight to be off continuously.
Press
to set the Backlight to turn on automatically whenever a key is pressed. It will turn off
automatically if no key has been pressed for 5 minutes.