
20
4.4 Conductivity Calibration
Before attempting a Conductivity calibration, ensure that the
smart
CHEM
-D
has been set up
correctly according to sections 4.1 to 4.3.
Automatic Temperature Compensation is done via a temperature sensor inside the Conductivity
sensor. It is therefore not necessary to calibrate the Temperature readout of the
smart
CHEM
-D
before taking Conductivity measurements.
1. Plug the Conductivity sensor into the
Cond / TDS
socket.
2. Rinse the Conductivity sensor in distilled water. Shake off as much water as possible. Blot the
outside of the sensor dry. DO NOT BLOT THE ELECTRODE PLATES.
Zero Calibration
3. Let the sensor dry in air.
4. Select Conductivity Calibration (
→
→
F1:Calibrate
→
→
F3:Cond.
).
5. The
smart
CHEM
-D
will recognise the low conductivity signal and attempt a Zero calibration.
For example…
Cond. 0*01uS
ZERO Calibration
Press Enter
6. Press
to calibrate. The Automatic Stability Function will now show a and highlight the
reading while it is unstable…
Cond.
0*01
uS
ZERO Calibration
Waiting to Stabilise
When the Conductivity reading has stabilised, the unit will calibrate itself.
To calibrate immediately without waiting for complete stability, press
to disable the
Automatic Stability Function.
The “
∗
” will not be removed after a zero calibration.
Continued over the page…
Summary of Contents for smartCHEM-D
Page 60: ...60 ...