25
Standard Calibration
7. Place the TDS sensor into a sample of TDS standard. Ensure that it is immersed correctly at
least to the top of the vent hole in the side of the sensor.
DO NOT
place the sensor directly into the bottle of standard. Discard the used sample of
standard after use.
Select TDS Calibration (
→
→
F1:Calibrate
→
→
F3:TDS
). The calibration screen
will be displayed with the TDS standard to be used. For example…
TDS 2*10ppK
2000.ppM Calibration
Press Enter
8. Press
to calibrate. The Automatic Stability Function will now show a and highlight the
reading while it is unstable…
TDS
2*10
ppK
2000.ppM Calibration
Waiting to Stabilise
When the TDS reading has stabilised, the unit will calibrate itself.
To calibrate immediately without waiting for complete stability, press
to disable the
Automatic Stability Function.
The “
∗
” will now be replaced by a decimal point if calibration was successful.
9. The
smart
CHEM
-D
is now calibrated for TDS and is ready for use in this mode.