1.9 Specifications
1.9.1 Conductivity
Ranges
Resolution
Accuracy
k=0.1 Sensor
0 to 2.000 µS/cm
0 to 20.00 µS/cm
0 to 200.0 µS/cm
0 to 2000 µS/cm
0 to 20.00 mS/cm
0.001 µS/cm
0.01 µS/cm
0.1 µS/cm
1 µS/cm
0.01 mS/cm
±0.2% of full scale of
selected range at 25
O
C
k=1.0 Sensor
0 to 20.00 µS/cm
0 to 200.0 µS/cm
0 to 2000 µS/cm
0 to 20.00 mS/cm
0 to 200.0 mS/cm
0.01 µS/cm
0.1 µS/cm
1 µS/cm
0.01 mS/cm
0.1 mS/cm
±0.2% of full scale of
selected range at 25
O
C
k=10 Sensor
0 to 200.0 µS/cm
0 to 2000 µS/cm
0 to 20.00 mS/cm
0 to 200.0 mS/cm
0 to 2000 mS/cm
0.1 µS/cm
1 µS/cm
0.01 mS/cm
0.1 mS/cm
1 mS/cm
±0.2% of full scale of
selected range at 25
O
C
Notes : Ranges are automatically selected. Exact auto-ranging points and full scales are subject to
sensor performance.
All ranges listed can be displayed in equivalent Siemens/metre units.
Sensor Type
.......................................... Glass body with two platinised platinum plates.
In-built ATC.
Temperature Compensation
................ Automatic, 0 to 100
O
C.
ATC co-efficient can be set from 0 to 6.00% /
O
C
individually for standard and sample.
ATC reference temperature is user-selectable for 25.0
O
C
or 20.0
O
C.
Calibration
............................................ Automatic zero and span calibration. Any calibration
standard from 20 µS/cm to 2000 mS/cm can be user set.
Sensor Span Range
............................... k=0.1 : k=0.075 to k=0.133
k=1.0 : k=0.75 to k=1.33
k=10 : k=7.5 to k=13.3