10
1.9.3 Resistivity
Ranges
Resolution
Accuracy
k=0.1 Sensor
0.50 to 99.00 M
Ω
.cm
50 to 9999 K
Ω
.cm
5.0 to 999.9 K
Ω
.cm
0.50 to 99.99 K
Ω
.cm
50 to 9999
Ω
.cm
0.01 M
Ω
.cm
1 K
Ω
.cm
0.1 K
Ω
.cm
0.01 K
Ω
.cm
1
Ω
.cm
±0.5% of full scale of
selected range at 25
O
C
k=1.0 Sensor
50 to 9999 K
Ω
.cm
5.0 to 999.9 K
Ω
.cm
0.50 to 99.99 K
Ω
.cm
50 to 9999
Ω
.cm
5.0 to 999.9
Ω
.cm
1 K
Ω
.cm
0.1 K
Ω
.cm
0.01 K
Ω
.cm
1
Ω
.cm
0.1
Ω
.cm
±0.5% of full scale of
selected range at 25
O
C
k=10 Sensor
5.0 to 999.9 K
Ω
.cm
0.50 to 99.99 K
Ω
.cm
50 to 9999
Ω
.cm
5.0 to 999.9
Ω
.cm
0.50 to 99.99
Ω
.cm
0.1 K
Ω
.cm
0.01 K
Ω
.cm
1
Ω
.cm
0.1
Ω
.cm
0.01
Ω
.cm
±0.5% of full scale of
selected range at 25
O
C
Note : Ranges are automatically selected. Exact auto-ranging points and full scales are subject to
sensor performance.
Sensor Type
.......................................... Glass body with two platinised platinum plates.
In-built ATC.
Temperature Compensation
................ Automatic, 0 to 100
O
C.
ATC co-efficient can be set from 0 to 6.00% /
O
C
individually for standard and sample.
ATC reference temperature is user-selectable for 25.0
O
C
or 20.0
O
C.
Calibration
............................................ Automatic zero and span calibration. Any calibration
standard from 20 ppM to 500 ppK can be user set.
Sensor Span Range
............................... k=0.1 : k=0.075 to k=0.133
k=1.0 : k=0.75 to k=1.33
k=10 : k=7.5 to k=13.3