38
7.2 Temperature Calibration Notes
1. Temperature calibration information is retained in memory when the
smart
CHEM
-C
is
switched off, even when the power supply is removed. This information can be recalled later
using the GLP function (see section 10).
2. Temperature does not need to be re-calibrated unless the Temperature or Conductivity/TDS
sensor is replaced or the meter is initialised.
3. The
smart
CHEM
-C
keeps the Temperature calibration data for the Temperature and k=0.1,
k=1.0 and k=10 Conductivity/TDS sensors separately so these can be swapped as required
without recalibrating Temperature.
7.3 Calibration Messages
1. If a temperature calibration has been successfully performed, the
smart
CHEM
-C
will display
the following message and the offset of the sensor.
Calibration OK
Offset=0.1
O
C
2. If a temperature calibration has failed, the
smart
CHEM
-C
will display the following message
and the failed offset value of the sensor.
Calibration Failed
Offset=11.0
O
C
3. The
smart
CHEM
-C
has an allowable Offset range of -10.0 to +10.0
O
C
. If calibration fails due
to the Offset being outside these limits, please consult the Troubleshooting guide (section 18.4)
for possible remedies.