6.2 Setting the Ratio calibration standard
This section is only applicable to Ratio*Standard mode. The factory default for this item is
2.76mS/cm. If this is satisfactory, go directly to section 6.3.
1. Select the Ratio Standard entry
(
→
→
F4:Setup
→
→
F1:Standards
→
→
F5:Ratio
).
The following screen is now displayed…
Ratio
Standard:2760.uS/cm
Range
20uS/cm to 2000mS/cm
2. Type in the value of the Ratio standard that is to be used for calibration, including the decimal
point. Use the
key to make any corrections.
3. Press
to save the value of the standard solution.
Alternatively, press
to quit without changing the current setting.
4. The
smart
CHEM
-C
will now ask you to enter the units for the Ratio standard…
Ratio
Standard:2760
Select Units
F1:uS/cm F2:mS/cm
(a) If the
smart
CHEM
-C
is set up for S/cm (section 3.2)…
Press
to set the Conductivity Standard as µS/cm.
Press
to set the Conductivity Standard as mS/cm.
or (b) If the
smart
CHEM
-C
is set up for S/m (section 3.2)…
Press
to set the Conductivity Standard as µS/m.
Press
to set the Conductivity Standard as mS/m.
Press
to set the Conductivity Standard as S/m.
or (c) If the
smart
CHEM
-C
is set up for dS/m (section 3.2), dS/m is set as the unit for the
standard.
5. The Ratio standard is now programmed for use at calibration for Ratio*Standard mode.