39
8. Auto Stability Function (ASF)
ASF adds an extra level of versatility to the
smart
CHEM
-C
. When ASF is activated, the
smart
CHEM
-C
monitors all parameters that are currently in use. When
ALL
parameters become
stable, the readings are frozen on the display.
ASF can be used in the following ways…
8.1 ASF During Normal Measurement
1. Press
in normal measurement mode. All active parameters are highlighted and the symbol
appears in the display. For example…
2
.
76
mS/cm
25
.
0
O
C
31/12/03
12:00:00
As each parameter becomes stable, the highlighting is removed. If any parameter subsequently
becomes unstable, the highlighting is applied again.
The highlighting is provided to give the operator a visual indication of exactly which
parameters have not stabilised in case user intervention is required.
2. When ALL parameters have stabilised, all highlighting will be removed and the will change
to a . All readings are now frozen. For example…
2
.
76
mS/cm
25
.
0
O
C
31/12/03
12:00:00
3. The operator can now make a note of the reading as required.
4.
To unfreeze the display in order to take the next reading, press
again to re-start the
ASF sampling process.
5. Press
a second time, while the symbol is being displayed, to turn ASF off.
8.2 ASF During Calibration
The Automatic Stability Function is automatically invoked during calibration. The stability
parameters programmed into the
smart
CHEM
-C
have a finer tolerance during calibration than
during normal measurement. This is done to ensure the most accurate possible calibration results.
Press
while the symbol is being displayed during calibration to turn ASF off and calibrate
immediately, before the reading has stabilised.
See the calibration sections of this manual for further details on the ASF function during
calibration.