44
10.2 Failed Calibration
If calibration has failed, the GLP function will reset the date and time for the failed parameter to
zero. The
smart
CHEM
-C
still shows the results for the last successful calibration, as shown in the
following example where the last Conductivity and TDS calibrations have failed…
k=1 Cell
Cond Zero=0.00 uS/cm 00/00/00 00:00
Cond k=1.10 @ 2760.uS/cm 00/00/00 00:00
Cond UnCalibrated
TDS Zero=0.00 ppM 00/00/00 00:00
TDS k=1.10 @ 2000ppM 00/00/00 00:00
TDS UnCalibrated F2:Back F4:Next
10.3 Printing GLP Information to the RS232 Port
The GLP information stored in the instrument’s memory can be sent to a printer or PC via the
RS232 port.
1. Switch the meter on.
2. Connect one end of the RS232 cable to the
Power/RS232
socket of the
smart
CHEM
-C
.
3. Connect the other end of the RS232 cable to an RS232 Printer, or to a Serial port on a PC.
4. Send the GLP information to the RS232 port:
→
→
F5:GLP
→
→
F3:Print
The message “
Printing GLP Data
” is displayed while sending the data to the RS232 port.
5. The GLP information is sent to the RS232 port in formatted ASCII text. For example…
smartCHEM-C V1.0 S1234 @ 31/12/2003 16:00
k=.1 Cond Zero= 0.001uS/cm @ 31/12/2003 12:00
k=.1 Cond k=0.11 @ 2760.uS/cm @ 31/12/2003 12:10
k=.1 TDS Zero= 0.000ppM @ 31/12/2003 12:00
k=.1 TDS k=0.11 @ 2000.ppM @ 31/12/2003 12:20
k=.1 Res. Zero= 0.00MOhms @ 31/12/2003 12:00
k=.1 Res. k=0.11 @ 362.3_Ohms @ 31/12/2003 12:30
k=.1 Temp. Offset= 2.0oC @ 31/12/2003 12:40
k=1 Cond Zero= 0.00uS/cm @ 31/12/2003 13:00
k=1 Cond k=1.10 @ 2760.uS/cm @ 31/12/2003 13:10
k=1 TDS Zero= 0.00ppM @ 31/12/2003 13:00
k=1 TDS k=1.10 @ 2000.ppM @ 31/12/2003 13:20
k=1 Res. Zero= 0.KOhms @ 31/12/2003 13:00
k=1 Res. k=1.10 @ 362.3_Ohms @ 31/12/2003 13:30
k=1 Temp. Offset= 3.0oC @ 31/12/2003 13:40
k=10 Cond Zero= 0.0uS/cm @ 31/12/2003 14:00
k=10 Cond k=10.1 @ 2760.uS/cm @ 31/12/2003 14:10
k=10 TDS Zero= 0.0ppM @ 31/12/2003 14:00
k=10 TDS k=10.1 @ 1452.ppM @ 31/12/2003 14:20
k=10 Res. Zero= 0.0KOhms @ 31/12/2003 14:00
k=10 Res. k=10.1 @ 362.3_Ohms @ 31/12/2003 14:30
k=10 Temp. Offset= 4.0oC @ 31/12/2003 14:40
Ratio Delta Span= 1.100 @ 31/12/2003 15:00
Temperature Offset= 1.0oC @ 31/12/2003 11:00
Ends