Topcon UA-10 Series Instruction Manual Download Page 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Instruction Manual 

 

 

Luminance & Chromaticity Uniformity Analyzer 

 

 

 

 

 

2D Luminance Colorimeter 

 

 

 

Rev2.2 

Rev. 2.0

 

Summary of Contents for UA-10 Series

Page 1: ...Instruction Manual Luminance Chromaticity Uniformity Analyzer 2D Luminance Colorimeter Rev2 2 Rev 2 0...

Page 2: ......

Page 3: ...D in TV accurately And it can measure interior panel switch in Automotive and illumination of LED OLED also And it is possible to easily correct the measurement data as compared with the measurement v...

Page 4: ...you or others This Caution mark indicates that ignoring this indication and mishandling the device may cause injury to you or others 1 or cause property damage 2 1 Injury stated here indicates the inj...

Page 5: ...tric shock Never disassemble or modify the device This may cause the fire and electric shock Never disassemble AC adapter Be sure to remove the dust or moisture around the outlet This may cause the fi...

Page 6: ...r insert the plug by wet hand This may cause you electric shock Never block the ventilation slot Ensure that the ventilation slot remain unobstructed or it may be cause of fire Use only specified scre...

Page 7: ...temperature tends to vary rapidly Never use or store the device in such place as is subject to heavy shock like falling or as tends to vibrate at any time Such places may damage this device equipped...

Page 8: ...the following matters can be performed by the user for maintenance Cleaning Lens For the dirt of lens please remove it with a soft cloth with diluted mild detergent saturated and then wipe it away wit...

Page 9: ...ling Driver 33 1 5 4 Connection Device to PC 37 1 6 Configuration in Windows 39 1 6 1 Screen Settings 39 1 6 2 Jumbo Frame Settings 39 1 6 3 Power saving settings 39 1 7 Part Names and Functions in Wi...

Page 10: ...Condition 2 4 UA 200A 82 3 7 1 Filter 82 3 7 2 Integral Time ND Filter Gain 83 3 7 3 Frequency 85 3 7 4 Specifying pick up area for optimizing measurement condition 86 3 7 5 Measurement Mode 87 3 8 S...

Page 11: ...151 3 13 3 Edit existing file 152 3 14 White Board Data Setting 153 3 14 1 Outline 153 3 14 2 White Board Measurement 154 3 14 3 Save White Board Data File 155 3 14 4 Open White Board Data File 156 3...

Page 12: ...ular Marker 211 5 1 10 Specifying optimization area by Mouse 212 5 1 11 Specifying optimization area by direct input 214 5 1 12 Release optimization area 216 5 1 13 Focus Adjustment 217 5 1 14 Magnifi...

Page 13: ...d Spot List 292 5 4 Split Spot View Operation 294 5 4 1 Open Split Spot View 295 5 4 2 Select Measurement Spot 296 5 4 3 Select All Measurement Spots 297 5 4 4 Deselect Measurement Spot 299 5 4 5 Dese...

Page 14: ...agram 379 5 8 6 Cancel Zoom in of Chromaticity Diagram 381 5 8 7 Turn ON Color Display Within Chromaticity Diagram 383 5 8 8 Turn OFF Color Display Within Chromaticity Diagram 385 5 8 9 Set Judgment A...

Page 15: ...ment condition File 441 5 16 6 Save Judgment condition File 442 5 17 Time series Measurement View Operation 443 5 17 1 Switch Time series View Display 443 5 17 2 Display Time series Graph 445 5 17 3 D...

Page 16: ...cipe of Currently Displayed Measurement Image 491 7 Display Window Operation 492 7 1 Tile 492 7 2 Close All 494 8 Help Operation 496 8 1 Topic Search 496 8 1 1 Open the Instruction Manual from HELP Me...

Page 17: ...keyboard This shows the reference section within the Manual This shows the reference document Note This explains what you should know or consider before starting the operation Memo This explains the r...

Page 18: ...es 1 Main body of UA 10 with object lens cap 1 2 AC Adapter 1 3 LAN cable CAT7 3m 1 4 Universal adapter 4 varieties of socket 1 Universal adapter may not be included in the package depending on the de...

Page 19: ...17 14 5 14 5 22 12 1 4 20 UNC NUT 1004 5 6 Screw 2x5 FW x4 Screw 2x10 FW x4 7 8...

Page 20: ...AWS Attachment lens AL UA4 1 UA 200AT only 2 AC Adapter 1 3 LAN cable CAT7 3m 1 4 CD ROM Installation program 1 Instruction manual PDF This document Application program Software Development Kit SDK Pa...

Page 21: ...cus ring Name Function Power connector Connects AC adapter DC12V Status LED Indicates status of the device Light off No power or No signal Green light Under communication or process Red blinking Error...

Page 22: ...Attach the spacer to the UA 10 and then attach the adapter for tripod screw over the spacer Fix together with the adapter and spacer by screws Tool screw Used to mount the UA 10 to fixture or stand M3...

Page 23: ...t Name Function Power connector Connects AC adapter Output plug DC12V Power switch Power ON OFF LAN Connector Connects LAN cable Use Straight through LAN cable Category 5e 1000BASE T TX or grater Hand...

Page 24: ...UA 200 series main body when the UA 200 series is mounted to a tripod Tripod screw 1 4 20 UNC NUT Depth 6mm Jig attachment screw Used to mount the UA 10 to fixture or stand M4 0 5 4mm Pitch 0 5mm Trip...

Page 25: ...ta It is not required to be connected to the photo detector It enables you to analyze and evaluate the data of the measurement image files at any place Note The following matters cannot be performed b...

Page 26: ...Intel CoreTM i5 4 core 2 8GHz or higher Memory 4GB or higher HDD 1GB or higher LAN Port 1000BASE T TX Gigabit Ethernet 1 port for color correction by using reference instrument RS 232C serial or USB 1...

Page 27: ...10 series and UA 200 series this program can be used as viewer program 2 Drivers You can install the driver for UA 10 series and UA 200 series In addition UA 200 and UA 200A has a different driver Dr...

Page 28: ...annot install the software without using the administrative right given account To install this software be sure to make the OS in the latest condition once using the Windows Update before installatio...

Page 29: ...Panel Add Remove Applications may not remove the software properly Transfer the contents of the CD ROM to USB memory please do never install May not install correctly Although Found New Hardware Wizar...

Page 30: ...e and right click to open the popup menu and then select Run as administrator When you install the application software to Windows7 operating system the User Account Control dialog may appear In that...

Page 31: ...e with the terms written here select the I accept the terms in the license agreement and click the Next If you don t accept the terms you cannot continue the installation work License Agreement Dialog...

Page 32: ...30 Dialog for installation running 6 Following dialog will appear Click Close to complete newly installation Dialog for complete installation...

Page 33: ...all of the Application software is as follows Note Be sure to close the application software before performing the uninstallation 1 Click the Application and uninstall starts Dialog for select install...

Page 34: ...32 Dialog for uninstall process 3 Following dialog will appear Click the Close to complete uninstall Dialog for uninstall complete...

Page 35: ...and the driver is installed with CD ROM version 4 0 0 or earlier driver update is required The measurement program may not operate properly To uninstall the driver using CD ROM version 4 0 0 or earli...

Page 36: ...bit in accordance with OS in the PC Click the OK to start installation Dialog for OS select Following dialogue will appear after complete installation It may take about ten seconds or longer to comple...

Page 37: ...th the UA 10 series and UA 200 series connected it will not be uninstalled normally 1 Insert the UA 10_200 Installation CD ROM into the CD drive Click the Driver button in following dialog Click the t...

Page 38: ...36 Following dialogue will appear after complete installation It may take about ten seconds or longer to complete the installation Wait for completing installation without any operation Memo...

Page 39: ...Never pull out or insert the plug by wet hand This may cause you electric shock UA 10 Series 1 Make sure that the plug of AC adapter is not inserted into an outlet 2 Connect the connector of AC adapt...

Page 40: ...AC adapter is not inserted into an outlet 2 Connect the plug of AC adapter to power connector of the device 3 Connect the LAN cable to LAN connector in the UA 200 and PC 4 Insert the Plug of AC adapte...

Page 41: ...her 5 Click Settings tab Set the color as follows Color quality Highest 32 bit 6 Set the DPI DPI Normal size 96 DPI How to select the control panel depends on the OS 1 6 2 Jumbo Frame Settings Jumbo f...

Page 42: ...40 1 7 Part Names and Functions in Window Names and functions of the parts displayed on the window are as follows 1 Menu bar 2 Toolbar 3 Rebar 4 Status bar...

Page 43: ...e currently displayed measurement image 6 3 1 Save Measurement Image Save All Measurement Images This function saves all retained measurement images to be saved 6 3 2 Save All Measurement Images Exit...

Page 44: ...his function enables you to freely arrange and measure up to 25 measurement spots 5 5 Random Spot View Operation Contour View This function classifies the measurement data and connects the same classi...

Page 45: ...nt date time in list form 5 11 Thumbnail View Operation RGB Color View This function arranges and displays RGB by using the measurement data 5 12 RGB Color View Operation Judgment result View This fun...

Page 46: ...rcibly interrupts the currently performed Normal Measurement or Time series Measurement 4 4 Cancel Measurement Setting Menu Recipe Setting This function performs all settings related to the measuremen...

Page 47: ...rrently displayed windows to be displayed by classifying them into four blocks top left top right bottom left and bottom right 7 1 Tile Close All This function closes all the currently displayed windo...

Page 48: ...are as follows Icon Menu Function File Open Measurement Image Save All Measurement Images Measurement Normal Measurement Time series Measurement Measurement Cancel View Live View Pseudo Color View St...

Page 49: ...47 RGB Color View L a b View Hue Chroma View Color System Pseudo Color View Judgment result View...

Page 50: ...uration icon will appear The procedure to check where saturation happen is as follows 1 Select the Tristimulus Value in the Pseudo Color View Pop up menu 2 Check the place where the saturation occurre...

Page 51: ...played measurement image is displayed The display format is as follows Year Month Date Hour Minute Second millisecond Area Measurable maximum area for the displayed measurement image is displayed in u...

Page 52: ...50 Measurement Time The measurement time of the displayed measurement image is displayed in ms Note Common Setting Status bar Setting displays the measurement image information...

Page 53: ...lp Topic Search Display corresponding help F5 Measurement Calculate optimal value of measurement condition F6 Normal Measurement F7 Time series Measurement F8 Setting Recipe Setting F9 Color Correctio...

Page 54: ...Install the Device Install the Detector 2 3 Installation Start the Software Start and Exit the Software 2 2 Start and Exit the Software Setup of the Measurement Conditions Recipe Setting 3 Recipe Sett...

Page 55: ...mode 1 Connect the device and the PC where the UA 10 UA 200 and UA 200A have been installed with LAN cables 2 Insert the plug of the AC cable into the outlet and turn ON the detector 3 From the Start...

Page 56: ...are as follows 1 When you start the Standard mode without connection to the device or turning off of the device the following dialog will appear Click OK 2 The following dialog appears Click the Yes t...

Page 57: ...software 2 If the measurement image to be saved is included in the saved measurement images the following dialog will appear Selecting the Yes saves all the measurement images to be saved and exit the...

Page 58: ...56 2 3 Installation 2 3 1 Live View 1 Launch the computer and start the application 2 From the Menu bar select the View Live View sequentially Or click icon on the Menu bar 3 The Live View will appear...

Page 59: ...w rotate the Focus ring to focus on the measurement target Note When you focus on the measurement target Striped patterns called moire sometimes appear on the measurement image The interference betwee...

Page 60: ...58 Note Use only specified screws when using the tripod screw and screw holes for jig attachment Do not tighten the screws any more than necessary Doing so might cause internal breakage...

Page 61: ...59 3 Setting 3 1 Open the Recipe Setting To open the Recipe Setting go through the following steps 1 From the Menu bar select Setting Recipe Setting sequentially Or press the F8 key...

Page 62: ...indow Cancel Disables the setting and closes this window Apply Enables the setting Enables you to continue the setting without closing the window Help Displays the Instruction Manual for the appropria...

Page 63: ...o through the following steps You can edit the active recipe file 1 Open the Recipe Setting dialog and click Select Recipe 2 Click the Open button and the Open File dialog will appear Select a recipe...

Page 64: ...be used in SDK 1 Open the Recipe Setting dialog 2 Edit the File name and Comment under Recipe file currently being displayed is applied and click the Save button to save the recipe file Click the OK b...

Page 65: ...le Sets field angle of objective lens The setting procedure is as follows Open Recipe Setting dialog and select Measurement Conditions 1 4 Select Wide or Standard from the Feld angle pull down menu in...

Page 66: ...ent Condition 1 4 Enter the values of measurement distance in the Measurement Distance edit box The measurement distance is from tip of the objective lens unit of the device to the measurement target...

Page 67: ...pin button Valid range of the number of times of the measurement is 1 999 When the Conduct optimization at the first measurement only check box is on the software optimizes a measuring setting at only...

Page 68: ...t time in the interval measurement exceeds the interval time When this item is checked If the measurement time exceeds the next interval measurement starting time the measurement is canceled at the ne...

Page 69: ...eraging is calculated automatically by integral time When you select the Manual the number of measurement count for the averaging can be specified When the Manual is selected the edit box become activ...

Page 70: ...s field angle of objective lens The setting procedure is as follows Open Recipe Setting dialog and select Measurement Conditions 1 4 Select Wide or Standard from the Feld angle pull down menu in the O...

Page 71: ...n 1 4 Set ON for the Attachment ND Filter is used check box of Attachment ND Filter Select the type of the attachment ND filter from the pull down menu Note For selection of attachment ND filter only...

Page 72: ...ent Condition 1 4 Enter the values of measurement distance in the Measurement Distance edit box The measurement distance is from tip of the objective lens unit of the device to the measurement target...

Page 73: ...pin button Valid range of the number of times of the measurement is 1 999 When the Conduct optimization at the first measurement only check box is on the software optimizes a measuring setting at only...

Page 74: ...t time in the interval measurement exceeds the interval time When this item is checked If the measurement time exceeds the next interval measurement starting time the measurement is canceled at the ne...

Page 75: ...the averaging is calculated automatically by integral time When you select the Manual the number of measurement count for the averaging can be specified When the Manual is selected the edit box becom...

Page 76: ...specify the integral time manually To set the integral time go through the following steps 1 Open the Recipe Setting dialog and click Measurement Condition 2 4 2 Select the Auto or the Manual in the...

Page 77: ...setting dialog and select the Measurement condition 2 4 2 Check the check box of Calculate the integration time from the frequency and the Frequency edit box will become active Specify the value of f...

Page 78: ...asurement condition 2 4 2 When Specify Optimization Area check box is On the setting in Optimization area in Live View is enabled Optimum measuring setting is calculated by using measured data in the...

Page 79: ...alog and select Measurement Conditions 2 4 2 Select one filter from Filer pull down menu Select one setting from Luminance Chromaticity Luminance and User specified In Luminance Chromaticity and Lumin...

Page 80: ...bination of Integral time ND filter and Gain in Auto In Manual setting of integral time ND filter Gain each can be specified manually The procedure of setting integral time ND filter and Gain is as fo...

Page 81: ...4 When Gain is fixed is checked software determine an optimum integral time under specified Gain Optimum integral time ND filter and Gain can be determined in advance of measuring Select Measure Calcu...

Page 82: ...setting dialog and select the Measurement condition 2 4 2 Check the check box of Calculate the integration time from the frequency and the Frequency edit box will become active Specify the value of f...

Page 83: ...select the Measurement condition 2 4 2 When the Specify Optimization Area check box is On the Optimization area setting in Live View is enabled Software determines an optimum measuring setting at spe...

Page 84: ...ialog and select Measurement Conditions 2 4 2 Select one filter from Filer pull down menu Select one setting from Luminance Chromaticity Luminance and User specified In Luminance Chromaticity and Lumi...

Page 85: ...ination of Integral time ND filter and Gain in Auto In Manual setting of integral time ND filter Gain each can be specified manually The procedure of setting integral time ND filter and Gain is as fol...

Page 86: ...4 When Gain is fixed is checked software determine an optimum integral time under specified Gain Optimum integral time ND filter and Gain can be determined in advance of measuring Select Measure Calcu...

Page 87: ...setting dialog and select the Measurement condition 2 4 2 Check the check box of Calculate the integration time from the frequency and the Frequency edit box will become active Specify the value of f...

Page 88: ...select the Measurement condition 2 4 2 When the Specify Optimization Area check box is On the Optimization area setting in Live View is enabled Software determines an optimum measuring setting at spe...

Page 89: ...rement condition 2 4 2 When setting ON for the Enable Layer Measurement check box layer is measured with the optimal integral time When setting ON for the check box the edit box is valid and you can s...

Page 90: ...4 3 8 1 Saturation Detection Sets the operation when saturation is detected during the measurement To set the saturation detecting operation go through the following steps Open the Recipe Setting dial...

Page 91: ...ontinue the measurement Click the No to discontinue the measurement If you do not select the Cancel Measurement when Saturation is Detected even if saturation is detected the measurement continues to...

Page 92: ...ing dialog and click Measurement Condition 3 4 2 Check the check box to use the initial measurement data as the starting point of the Time series Measurement Display When this check box is checked the...

Page 93: ...save the measurement image go through the following steps 1 Open the Recipe Setting dialog and click Measurement Condition 3 4 2 Check Save Measurement Image Auto Save Measurement Image to save the m...

Page 94: ...through the following steps 1 Open the Recipe Setting dialog and select the Measurement Conditions 3 4 2 Diagonal Correction is On the Diagonal correction is enabled to measured image at each measuri...

Page 95: ...ng angle rotation is applied to the measurement image at every measurement 3 When setting ON for the Use Rotation Angle of Live Image check box the rotation in the angle for the live image is applied...

Page 96: ...ata This smoothing enables you to acquire stabilized measurement data 1 Open the Recipe Setting dialog and select Measurement Conditions 3 4 2 Select one smoothing type from Smoothing pull down menu H...

Page 97: ...wn menu of MURA Emphasis select the MURA Emphasis formula to be applied Max value division A max Divides the image which is the filtering result before MURA Emphasis processing by the maximum value of...

Page 98: ...number by the image which is the filtering result before MURA Emphasis processing Specified order division B A Divides the image which is the filtering result up to the order number specified by Divis...

Page 99: ...the following steps 11 Appendix Terminology Median filter 1 Open the Recipe Setting dialog and select Measurement Condition 4 4 2 From the Size pull down menu of Median parameters select the filter s...

Page 100: ...er 1 Open the Recipe Setting dialog and select Measurement Condition 4 4 2 From the Size pull down menu of Gaussian parameters select the filter size to be applied 3 Set the values for STDEV X and STD...

Page 101: ...igmaColor and SigmaSpace of Bilateral parameters The setting range is 0 to 999 for Diameter of each pixel neighborhood and 0 to 999 99 for SigmaColor and SigmaSpace As Diameter of each pixel neighborh...

Page 102: ...list and change the order with the UP DOWN buttons To delete the set item select the desired item from the Order Filter list Press the delete button and the selected item is deleted from the Order Fi...

Page 103: ...erform the one item processing 1 Open the Recipe Setting dialog and select Measurement Condition 4 4 2 When the Activate Filter Correction check box is ON Filter Correction is applied to the measureme...

Page 104: ...correction definition go through the following steps 1 Open the Recipe Setting dialog and click Color Correction 2 Select whether or not the color correction factor to be enabled The checked color cor...

Page 105: ...nabled from the list The checked color correction factors become enabled You can select several color correction factors and multiply the selected color correction factors For example when you select...

Page 106: ...r correction factors or no one When you select several color correction factors the factors are multiplied You can select up 20sets of color correction factors In UA 200 series 2sets of correction fac...

Page 107: ...1 Open the Recipe Setting dialog and click Spot Correction 2 Select whether or not the spot correction factor become enabled The checked spot correction factor will becomes enabled You cannot select t...

Page 108: ...en button to open File open dialogue Select an area correction file to be loaded Loaded File is enabled When Enable Area Correction is On the Area correction in Tristimulus values XYZ are multiplied a...

Page 109: ...o select and set the white board correction go through the following steps 1 Open the Recipe Setting dialog and select White Board Correction 2 Select whether or not the white board correction factor...

Page 110: ...wing steps 1 Open the Recipe Setting dialog and click Define Standard White Point Select the kind of standard light used as the standard white point from pull down menu in the Standard light 2 When ot...

Page 111: ...e CSV File Type Pull down menu select the view for which to set the Save CSV function All Data These data are saved on the Pseudo Color View and RGB Color View The values of the measurement image in e...

Page 112: ...This data is saved on the Cross section Judgment result View In the standard spot and the random spot maximum and minimum values can be displayed In the case of All Data it is possible to output the...

Page 113: ...size to increase and requires a longer time Note Selecting the All Data may delay the display temporarily Outputting from the View Pop up menu will cause the following restrictions When outputting All...

Page 114: ...put the CSV file s during the measurement go through the following steps Note If there is no measurement data set in the selected output item no file is output If the software is executed in View mode...

Page 115: ...is checked the destination folder path is displayed To change the path click Browse button The folder path is displayed in Explorer style Select the location in which to save the file and click OK but...

Page 116: ...alculates the correction factor so that the measurement data obtained by the standard unit and by the device are the same value Specify the pattern and the size of spot to coincide with measurement ar...

Page 117: ...ary using the existing correction factor causes a large error in the measurement data of the Standard Unit To ensure the reliability of the measurement data we recommend that you execute the Color Cor...

Page 118: ...tandard Unit go through the following steps 1 From the Menu bar click Setting Color Correction Wizard sequentially Or press the F9 key 2 The Color Correction Wizard appear Warm up the instrument for 5...

Page 119: ...or the Edit Color Correction Factor and click Next The Calculate Color Correction Factors Determining the color correction factor by using the device and the Standard Unit The Edit Color Correction De...

Page 120: ...lign the center of the measurement target and the center marker of the device Click the Next after aligning them Markers Select the cross marker or the diagonal marker as the center marker to be displ...

Page 121: ...ttern from the round or square Spot Size This function sets the length of the diameter or side line of the measurement spot Setting range 0 01 500 mm Single color Multicolor Select the type of color c...

Page 122: ...lor After completing the measurement click Next Multicolor Performs measurement according to the number of colors When the measurement is finished the following message dialog box appears Change the o...

Page 123: ...sing the device is displayed If no problem is found in the measurement result click Next The measurement result of a point which could not be measured is displayed as In the case of two correction fac...

Page 124: ...ear Select how to calculate color correction factors Calculation by Standard Unit Retrieves the measurement data from the Standard Unit directly and calculate the correction factor Pressing Next moves...

Page 125: ...g box for values of luminance and chromaticity will be active In the case of Multicolor it is necessary to set the values for each of the selected colors Luminance Valid range 0 005 9 999 999 Chromati...

Page 126: ...nication interface Standard Unit Select the Standard Unit model from the Pull down menu Note Be sure to select the Standard Unit connected to the PC Each Standard Unit has its own communication protoc...

Page 127: ...interface actually connected Connection Test Perform the connection test with the Standard Unit If the connection test is succeed the following dialog will appear Click Next button to starts the measu...

Page 128: ...R UL2 LEDW 38400 7 ODD 1 Normal Type BM 7A 38400 7 ODD 1 BM 7A BM 5A 1200 7 ODD 1 BM 5AS 38400 7 ODD 1 BM 5AS 10 Standard Unit Measurement Status will appear After completing the measurement Next beco...

Page 129: ...n factor is calculated based on the measurement result of Standard Unit After confirming it press Next UA 10 series Single color UA 200 series Single color Multicolor The color correction factor is no...

Page 130: ...12 Display Color Correction Factors will be displayed The color correction factor is calculated based on the measurement result of Standard Unit After confirming it press Next UA 10 series UA 200 ser...

Page 131: ...129 13 Click the Finish button To use the created color correction factor select it on the Recipe Setting Color Correction Page 3 10 2 Select Color Correction Definition...

Page 132: ...olor correction factor created before To change the color correction definition go through the following steps 1 From the Menu bar click Setting Color Correction Wizard sequentially Or press the F9 ke...

Page 133: ...ick Next 4 The color correction definition list will be displayed On this list you can edit the correction factor KX KY and KZ and the comment Click the cell to be edited and edit it When you finish t...

Page 134: ...132 UA 200 series 5 The Finish window is displayed Click Finish...

Page 135: ...ete any desired color correction factor definition go through the following steps 1 The Color Correction Definition List is displayed 2 Click the definition to be deleted on the list to highlight the...

Page 136: ...rrection factor definition go through the following steps 1 From the Menu bar click Setting Color Correction Wizard sequentially Or press the F9 key 2 The Color Correction Wizard appear Warm up the in...

Page 137: ...Spot Correction Factor is used to create the spot correction definition by using the detector and the standard spot data The Edit Spot Correction Definition is used to delete the spot correction defin...

Page 138: ...t Vertical Measurement Spot Count Select the quantity of the measurement spots in the vertical direction to one of 3 5 7 9 11 13 15 17 19 21 Horizontal Measurement Spot Count Select the quantity of th...

Page 139: ...evice After aligning them click Next Markers Select the cross marker or the diagonal marker as the center marker to be displayed Measurement Spot Set whether the set measurement spot is displayed or n...

Page 140: ...elect the spot standard file After selecting it click Next 8 The spot correction factor definition is displayed The result of the calculation of the correction factor based on the measurement data of...

Page 141: ...139 9 Click Finish To activate the created color correction select it on the Recipe Setting Spot Correction Page 3 10 3 Select Spot Correction Definition...

Page 142: ...Definition List To delete the spot correction factor definition go through the following steps 1 The Spot Correction Definition List is displayed 2 Right click the highlighted definition to be deleted...

Page 143: ...bar click Setting Common Setting sequentially Or press the F10 key 2 The Common Setting dialog will open All of the following buttons displayed on the Recipe Setting Dialog are the same function OK E...

Page 144: ...go through the following steps 1 Open the Common Setting dialog If the initial layout is not opened click Initial Layout Setting from the list on the left 2 On the four separate windows created by di...

Page 145: ...ed window are displayed as shown below after clicking Display Initial Layout with the Initial Layout Initial Layout Setting Displayed Windows Note Selecting any of various data sheets does not allow y...

Page 146: ...er integral time Sensitivity increase but responsibility decrease Specify the appropriately integral time depending on the brightness of the measurement target Valid range 0 03 1000 ms UA 10 series 0...

Page 147: ...Dialog Sets whether or not the Check Measurement Conditions dialog is displayed When this function is checked the Check Measurement Conditions dialog is displayed in every measurement When not checke...

Page 148: ...View When this function is checked Tooltip appear on the position where the mouse place on the Pseudo Color View When the checkbox is not checked the Tooltip is not displayed Tooltip is a small window...

Page 149: ...ch data sheet This setting affected the data sheet and the output CSV file Graphs and various property dialogs are not affected To perform the formatting go through the following steps Open the Common...

Page 150: ...the digits after the decimal point Therefore the number of digits after the decimal point set by Formatting may not be displayed depending on the measurement data Example When 6 digits are set in the...

Page 151: ...the status bar method go through the following steps 1 Open the Common Setting dialog Click the Status Bar Setting from the list on the left 2 Select items to be displayed in the status bar Please ref...

Page 152: ...This function can divide measure image in a grid shape and multiply correction factor of Tristimulus values to each grid In this menu you can create and edit an Area correction file Select the Settin...

Page 153: ...ontal split count and Vertical split count and then edit each correction factor in grid 4 Press Save button to save the Area correction file An Area correction file is saved as CSV format An Area corr...

Page 154: ...rrently being displayed is applied can be edited directly Set Tristimulus value Horizontal Split Count and Vertical Split Count And then edit each correction factor in grid 4 Click the Save button to...

Page 155: ...White Board Data Setting 3 14 1 Outline It is possible to set the white board data which will be the calibration standard required for the object color measurement Select the Setting White Board Data...

Page 156: ...he Measurement go through the following steps 1 From the Menu bar click Setting White Board Setting sequentially The White Board Setting dialog will open 2 Select the White Board Measurement 3 12 4 En...

Page 157: ...board data file To save the white board data file go through the following steps Open the White Board Setting dialog Edit the Comment and click the Save button to save the white board data file Select...

Page 158: ...he white board data file To open the white board data file go through the following steps 1 Open the White Board Setting dialog Click the Open button to open the white board data file Select Browse an...

Page 159: ...2 Measurement Image Load List dialog will open The measurement date time and comment of the loaded measurement image are displayed To open the file click OK 3 The loaded white board data file is open...

Page 160: ...at application startup the white board data file To Read automatically at application startup the white board data file go through the following steps Open the White Board Setting dialog Check the Rea...

Page 161: ...tilted image which is measured at tilting angle to a rectangular Sets a correction setting applies the setting to a measured image saves the correction factor data in a correction factor list and dele...

Page 162: ...in keyboard to roll back to one previous operation 1 Select the Setting Diagonal Correction when a measured image is displayed The Diagonal Correction view will appear When you select the Setting Diag...

Page 163: ...of an area to be corrected Select points in a clockwise direction from first point 3 Specify two opposite angle of rectangle to be corrected An outline of a rectangle connecting two specified opposite...

Page 164: ...162 4 Measured image in corrected rectangle shape will be displayed...

Page 165: ...urement Image 2 Diagonal Correction is applied to measured image on the display Once a diagonal correction is applied to a measured image a parameter of a diagonal correction is saved 3 15 5 Save Dale...

Page 166: ...e following steps 1 Right click on the Diagonal correction View and pop up menus appear Select the Deselect the Setting 2 The diagonal correction is canceled and the image return to original image Whe...

Page 167: ...a diagonal correction To save a parameter of a diagonal correction go through the following steps 1 Right click on the Diagonal correction view and pop up menu appear Select the Save to Settings List...

Page 168: ...r of a diagonal correction highlight a parameter to be deleted and right click on the row A pop up menu appear and select the Delete to delete the parameter of the diagonal correction Select one from...

Page 169: ...ays Display the Setting List To display Settings List go through the following steps 1 Right click on the Diagonal Correction view and pop up menu appear Select the Display the Settings List 2 The Dia...

Page 170: ...Select the Properties You can edit a property directly if you want to conduct fine adjustment of property Click the OK button to apply a property to an image in the Diagonal Correction view Coordinat...

Page 171: ...our Arithmetic Operations 3 16 1 Outline Performs four arithmetic operations between the measurement images Select Setting Four Arithmetic Operations under the condition that the measurement image is...

Page 172: ...operations will be performed The measurement image which should be the standard 2 Specify the standard measurement image from the measurement image list of the standard data at the right side on the s...

Page 173: ...he Select all button it is possible to select all of the measurement images to which four arithmetic operations can be applied from the measurement image list 4 Enter the offset value on Four Arithmet...

Page 174: ...FF Pseudo color setting of the arithmetic operations results are displayed by using the entered maximum and minimum values When Max value and Min value are displayed in red there is a value that is be...

Page 175: ...e Click the Adding measurement image button and Measurement Image List is displayed Select any of results of four arithmetic operations and right click Edit File Name or Comment Editing to change the...

Page 176: ...contents on the displayed list are added as a measurement image If you do not want to add them click the Cancel button The system returns to the Four Arithmetic Operations dialog box The Apply button...

Page 177: ...ing is performed for the measurement image Under the condition that the measurement image is displayed select Setting Filtering When the image is displayed with gray scale the visibility of MURA uneve...

Page 178: ...filtering result before MURA Emphasis processing by the maximum value of the image which is the filtering result up to the order number specified by Division order number Avg value division A avg Div...

Page 179: ...ion B A Divides the image which is the filtering result up to the order number specified by Division order number by the image which is the filtering result before MURA Emphasis processing Example In...

Page 180: ...rough the following steps 11 Appendix Terminology Median filter 1 Select Setting Filtering and the Filtering dialog will open 2 From the Size pull down menu of Median parameters select the filter size...

Page 181: ...filter 1 Select Setting Filtering and the Filtering dialog will open 2 From the Size pull down menu of Gaussian parameters select the filter size to be applied 3 Set the values for STDEV X and STDEV...

Page 182: ...aColor and SigmaSpace of Bilateral parameters The setting range is 0 to 999 for Diameter of each pixel neighborhood and 0 to 999 99 for SigmaColor and SigmaSpace As Diameter of each pixel neighborhood...

Page 183: ...delete the set item select the desired item from the Order Filter list Press the delete button and the selected item is deleted from the Order Filter list The set value when pressing the add button i...

Page 184: ...er and filter parameters 1 Select Setting Filtering and the Filtering dialog will open 2 In the measurement image list of Standard data place a check mark on the check box of the measurement image to...

Page 185: ...correction image is displayed on Result The filter correction image file is not created To create the filter correction image file execute 3 17 8 Adding Filter Correction Image to Measurement Image T...

Page 186: ...The MURA Emphasis unevenness emphasis image can be created by adjusting the processing order and filter parameters When Filtering execution has already been performed the Adding measurement image butt...

Page 187: ...surement image which is being displayed on Standard data is displayed and this filter correction image is added to the measurement image Initial setting of filter correction image File Name Year Month...

Page 188: ...e To reflect the filter correction conditions in Recipe go through the following steps 1 Select Setting Filtering and the Filtering dialog will open 2 After setting the filter correction conditions pr...

Page 189: ...layed or based on the Initial Layout or currently displayed layout To perform the Normal Measurement go through the following steps 1 From the Menu bar click Measurement and select the Normal Measurem...

Page 190: ...g Display Measurement Condition Confirmation Dialog for Every Measurement 3 12 4 Environment Setting 3 To start the measurement click Start When the color correction factor has been enabled the color...

Page 191: ...al values of integral time is determined in each measurements When you continuously measure the same kind of targets obtain optimum measurement condition by using the Calculate optimal values of measu...

Page 192: ...ut Setting When continuous measurement or interval measurement is performed the measurement image are overwritten by next measurement image The above is a case where the layout is set as follows Top l...

Page 193: ...ugh the following steps 1 From the Menu bar click Measurement and select Time series Measurement Or press the F7 key 2 Check Measurement Conditions will appear To check the content and start the measu...

Page 194: ...splayed after clicking Start Click Start to display the following Check Color Correction dialog To start the measurement click Start To cancel the measurement click Cancel UA 10 Single color UA 200 Si...

Page 195: ...urement Graph and Time series Measurement Data Sheet The layout can be changed after measurement but the default layout cannot be changed In the following cases a warning is displayed and the measurem...

Page 196: ...e will actually perform the automatic calculation of optimum measurement condition When the measurement targets have relatively small luminance variation this function enable you to shorten the measur...

Page 197: ...optimal value calculation UA 10 series UA 200 series However when the AUTO is selected in the Integral Time Setting in the Recipe the following confirmation dialog will appear When you click Yes the...

Page 198: ...the following steps 1 From the Menu bar click Measurement Cancel sequentially Or click the Cancel button in the Check Status dialog Or click the STOP button on the Menu bar 2 The following dialog is...

Page 199: ...the measurement accuracy and measurement time If Auto is selected in Measurement Condition 2 4 Integral Time the optimum integral time is calculated automatically before each measurements In the conti...

Page 200: ...easurement object in real time This function is used to confirm the position of the target and adjust the focus of the measurement object To open the Live View go through the following steps 1 From th...

Page 201: ...199 2 Live View is opened...

Page 202: ...o adjust the contrast go through the following steps 1 Right click on the Live view and the pop up menu will appear then select the Contrast 2 The scroll bar for contrast will appear Adjust contrast b...

Page 203: ...on setting Live setting is set to AUTO Depending on memory usage in PC an adjustment operation may not work on Live view immediately In this case Renew a Live view by selecting Stop and then select Pl...

Page 204: ...UA 10 series Switches color or gray scale in the Live view To switch color or gray scale in the Live view go through the following steps Right click on the Live view and the pop up menu will appear S...

Page 205: ...Live View Restarts the Live View from the pause state To play the Live View go through the following steps Right click anywhere within Live View screen pop up menu will appear Select Play to restart...

Page 206: ...ive image or save the snapshot To stop the Live View go through the following steps 1 Open the Live View Right click anywhere in Live View screen and Pop up menu will appear Select Stop 2 The Live Vie...

Page 207: ...To switch the markers go through the following steps Open the Live View Right click anywhere in the Live View screen Pop up menu will appear Select Switch Marker Select Cross line or Diagonal as the m...

Page 208: ...206...

Page 209: ...st the position of the measurement target To set the Rectangular marker by using mouse in the Live View the steps are as follows 1 The Rectangular marker area is set by dragging mouse on the Live View...

Page 210: ...208 Rectangular marker area is displayed...

Page 211: ...sition of the measurement target To set the Rectangular marker by entering values in the Live View the steps are as follows 1 Right Click on Live View Image and Pop up menu will appear Select Property...

Page 212: ...210 3 Rectangular marker will appear...

Page 213: ...lar marker To release the displayed Rectangular marker displayed on the Live View the steps are as follows 1 Right Click on Live View Image and Pop up menu will appear Select Rectangular Marker releas...

Page 214: ...pecify optimization area by using mouse go through the following steps 1 Click the start point of the optimization area in the Live view and drag the mouse over the area to end point to specify the op...

Page 215: ...213 3 Optimization area is displayed...

Page 216: ...To specify the optimization area by entering values go through the following steps 1 Right click the Live view and the pop up menu will appear And select the optimization area Property 2 The optimiza...

Page 217: ...215 3 Optimization area is displayed...

Page 218: ...rea Cancels the optimization area To cancel the optimization area go through the following steps 1 Right click on the Live view and the pop up menu will appear And select the optimization area Release...

Page 219: ...ocus in real time You can easily adjust the focus by referring to the score displaying on the Live view To adjust the focus go through the following steps 1 Right click on the Live View and pop up men...

Page 220: ...y not reach 100 depending on brightness and contrast of the target We recommend that pattern for focusing is displayed when focusing in order to adjust the focus easily Contrast in Live view is automa...

Page 221: ...u want to zoom in on a specific point of the live view To Magnification Display go through the following steps 1 Place the mouse to enlarge display on the Live View 2 Right click on the Live View and...

Page 222: ...220 3 Enlarged display window appears...

Page 223: ...221 4 Change enlargement position by clicking the mouse Rotating the mouse wheel while pressing the Ctrl key enables you to change the display magnification Memo...

Page 224: ...and the pop up menu is displayed Select Rotation 2 The Rotation dialog box is displayed Set ON for the Rotate Live Image check box Operate the scroll bar or enter a value to the angle edit box directl...

Page 225: ...in the Live View screen Pop up menu will appear Select Resolution Select High or Low as the resolution to be displayed When High is selected high resolution is displayed When Low is selected the low r...

Page 226: ...d to measurement data Slight difference in measurement data are color coded for finding easily uniformity and Mura on the target To open the Pseudo Color View go through the following steps 1 From the...

Page 227: ...me in Standard Spot View Split Spot View Random Spot View and Contour View Color system Pseudo Color View Judgment result View 1 Activate the Pseudo Color View and right click on the view Pop up menu...

Page 228: ...an be displayed vertically and horizontally using the scroll bar since the entire measurement image cannot be displayed at a window Move to the portion desired to be seen by using the scroll bar Rotat...

Page 229: ...s the same in Standard Spot View Split Spot View Random Spot View and Contour View Color system Pseudo Color View Judgment result View 1 Right click on the view and Pop up menu will appear Select the...

Page 230: ...Scale The color can be displayed by a gray scale or pseudo color In the Gray Scale display the frame border color of the measurement spot is entirely switched to red in the Random Spot View Standard...

Page 231: ...yed on the Pseudo view To change the item to be displayed go through the following steps This operation is the same in Standard Spot View Split Spot View Random Spot View Contour View and Cross Sectio...

Page 232: ...e Tristimulus value list Select the Tristimulus value to which the display is to be switched The measurement image on the view is switched to the selected item of the measurement image The selected it...

Page 233: ...city item is displayed To change the item to be displayed go through the following steps This operation is the same in Standard Spot View Split Spot View Random Spot View Contour View and Cross Sectio...

Page 234: ...e view is switched to the measurement image of the selected chromaticity The selected chromaticity is permanently retained and if the software is restarted the measurement image is displayed with the...

Page 235: ...is redrawn to be color coded The trimmed measurement image is affected to other views To perform the trimming by using mouse go through the following steps 1 Open the Pseudo Color View 2 Click the sta...

Page 236: ...image and the trimmed image This is because the maximum values and minimum values for the whole area are different from those of the trimmed image When the area is trimmed the display colors are allo...

Page 237: ...u Select Trimming Property from the Pop up menu 3 Trimming dialog is displayed Click OK after setting the start and end points of the Trimming By setting ON for the Save to All Applicable Measurement...

Page 238: ...for the whole area are different from those of the trimmed image When the area is trimmed the display colors are allotted in accordance with the measurement data within the trimmed area allowing you...

Page 239: ...ge here are two way to cancel the trimmingTo cancel trimming go through the following steps Cancel Trimming on Pseudo color View 1 From the Thumbnail View select the measurement image for which trimmi...

Page 240: ...38 3 The trimming is canceled and the original measurement image is displayed Cancel Trimming via pop up menu 1 From the Thumbnail View select the measurement image for which trimming is to be cancele...

Page 241: ...39 2 Right click on the view and The Pseudo Color View Pop up menu will appear From the Menu select Trimming OFF sequentially 3 The trimming is canceled and the original measurement image is displayed...

Page 242: ...lor bar on Pseudo Color View To set the Color bar the steps are as follows 1 Click on Pseudo Color View to activate Window 2 Right click on the Pseudo Color view and the Pop up menu will appear Select...

Page 243: ...romaticity u v color display 4 If Display Color bar OFF is selected from the Pop up menu the Color bar will be hidden Color bar are adjustable on the Pseudo Color View Adjustment in the color bar take...

Page 244: ...maximum and average values can be displayed on Pseudo Color View To view the minimum maximum and average values the steps are as follows 1 Click on Pseudo Color View to activate Window 2 Right click o...

Page 245: ...ax Green yellow Chromacities xy color Min White Max Black xy display Chromacities u v color Min White Max Black u v display Only when Tristimulus value Y is selected Uniformity is displayed below the...

Page 246: ...is operation is the same in Live View Standard Spot View Split Spot View Random Spot View Contour View Cross Section View Chromaticity Diagram View Histogram View and 3D View RGB Color View Color syst...

Page 247: ...245 3 The Explorer window opens Specify the path file name and file format to be saved BMP TIFF or PNG format are available...

Page 248: ...s the same in Standard Spot View Split Spot View Random Spot View RGB Color View Color system Pseudo Color View and Judgment result View Output items to be saved are specified on Recipe Setting CSV Se...

Page 249: ...Judgment result random spot rdmj_ Judgment result standard spot stdj_ Judgment result split spot mtxj_ Judgment result all pixel allj_ If the Pseudo Color View Save CSV is selected when the trimming...

Page 250: ...ure Deviation Excitation Purity and Dominant Wavelength are recorded sequentially In the Recipe Setting CSV Setting File Type All Data the unchecked items are not recorded When maximum minimum and ave...

Page 251: ...olor Temperature Deviation Excitation Purity and Dominant Wavelength are recorded sequentially On each measurement data the difference in the measured value between each spot and the center spot numbe...

Page 252: ...Dominant Wavelength are recorded sequentially The data is recorded according to the split count set on Property of Split Spot View The spot number is recorded in the first column of the first row loca...

Page 253: ...lor Temperature Deviation Excitation Purity and Dominant Wavelength are recorded sequentially On each measurement data the difference in the measured value between each spot and the center spot number...

Page 254: ...Chromaticity u and v Color Temperature Deviation Excitation Purity and Dominant Wavelength are recorded sequentially In each measurement data the measurement value of the intersection point the measu...

Page 255: ...minant Wavelength are recorded sequentially The measured data are arranged sequentially from the top down in line for the image measurement count from the measurement start time and the measurement da...

Page 256: ...recorded sequentially For the selected judgment items a blank is recorded for OK and NG is recorded for NG When all the selected items are OK the general OK NG judgment result in the spot is OK When o...

Page 257: ...al time When the Property dialog is closed after setting the display color the display color is immediately applied to the following views Standard Spot View Split Spot View Random Spot View Contour V...

Page 258: ...f the Tristimulus values X Y Z and the Chromaticity x y u v are displayed on the table on the right side dialog When the setting is completed click any button OK Enables the setting and closes this wi...

Page 259: ...nces in measurement image can be emphasized The valid range is from the minimum to the maximum value of the measurement data When changing measurement items of Tristimulus values and chromaticity the...

Page 260: ...maximum value range of measurement data 3 Pseudo Color Property will appear Set OFF for the Color scale synchronization check box Set the upper and lower limit by using the Set Upper Limit Lower Limit...

Page 261: ...for the Color scale synchronization check box Enter directly the value in the edit box beneath the Set Upper Limit Lower Limit slide bar When selecting ON for the Color scale synchronization check box...

Page 262: ...mit of color bar To specify the Color range go through the following steps 1 Open the Pseudo Color Property 2 Slide the Color Range slide bar while watching a variation of color in the Pseudo Color Vi...

Page 263: ...the image go through the following steps Masking is applied to the measurement image of the currently displayed 1 Open the Pseudo Color Property Adjust the Set Upper Limit Lower Limit slide bar while...

Page 264: ...t image where measurement values are within the upper and lower limit are displayed If you save the measured data in a CSV file with the masking On the measurement data with image displaying only are...

Page 265: ...t go through the following steps Locking the upper and lower limit is applied to the currently displayed measurement image 1 Open Pseudo Color Property Set the Set Upper Limit Lower Limit slide bar wh...

Page 266: ...ata and chromaticity data view without re designation To apply the range to all the Tristimulus values go through the following steps 1 Check the Mask checkbox Click Apply to check the mask setting ra...

Page 267: ...ower Limit Color Range and Lock Adjustment for all the setting values of Pseudo Color Adjustment to be applied 2 To apply the respective adjustment values of the pseudo color to all the loaded images...

Page 268: ...f the measurement data The lower limit value of the pseudo color adjustment is not changed However only the measurement data whose upper limit value is below the upper limit value of the pseudo color...

Page 269: ...urement image go through the following steps 1 Open Pseudo Color View 2 Pseudo Color View is opened Right click to open the pop up menu and select Rotation from the menu 3 When Rotation is selected th...

Page 270: ...k to close the Rotation dialog box By setting OFF for the Rotate Measurement Image check box you can return the measurement image to the status where it is not rotated By setting ON for the Save to Al...

Page 271: ...ea of 500 or more pixels per the measurement spot In the JEITA standard measurement method the vertical direction based measurement is specified However in the UA 10 measurement having an angle to the...

Page 272: ...Standard Spot View To open the Standard Spot View go through the following steps 1 From the Menu bar select View Standard Spot View sequentially Or click the icon on the Menu bar 2 The Standard Spot...

Page 273: ...tandard spot measurement data in a spreadsheet style To display the data sheet go through the following steps 1 Open the Standard Spot View 2 Right click anywhere within the Standard Spot View to open...

Page 274: ...Data Sheet is displayed 5 3 3 Switch Display Items of Standard Spot Data Sheet The measurement values which saturation occur in the Tristimulus value are displayed as over and the chromaticity is disp...

Page 275: ...of the data sheet in the standard spot view To switch the display items of the data sheet go through the following steps This operation is the same in the Random Spot Data Sheet 1 Open the Standard S...

Page 276: ...enu list is selected the item is newly added to the data sheet If the item checked in the Display Items menu list is selected the item is deleted from the display items displayed on the data sheet Sin...

Page 277: ...File Format Saves the content displayed on the data sheet in CSV file format This operation is the same as in the Random Spot Data Sheet 1 Open the Standard Spot Data Sheet 2 Right click anywhere with...

Page 278: ...plorer window will open Select the save destination path and specify the file name and format for the file to be saved The default file name is in the style of date time For the file format you can se...

Page 279: ...277 5 This is an image opened by Excel The content displayed on the data sheet is saved Standard Spot Data Sheet Window Standard Spot Data Sheet CSV File...

Page 280: ...sed on outer frame The setting is applied to the currently displayed measurement image and to all measurement images To open the Standard Spot Property go through the following steps 1 Open the Standa...

Page 281: ...ant to terminate the setting operation click one of the following buttons OK Enables the setting and closes this window Cancel Disables the setting and close this window Apply Enables the setting and...

Page 282: ...ot via standard spot list 5 3 15 Display Standard Spot List 1 Display the Standard Spot Property window 2 Select the spot pattern of the Initial Measurement Spot from the Pull down menu Round or squar...

Page 283: ...d split and the even split The difference between standard split and even split is as shown below For JEITA standard select Standard Split A For standard split For even split The measurement spot diff...

Page 284: ...rd from the Pull down menu 3 Select Measurement Spot Count from the Pull down menu For the spot count select one of four types 5 9 13 or 25 points 4 To specify the measurement from the outer frame of...

Page 285: ...pin buttons The setting ranges for vertical and horizontal dimension are from 1 to 999 mm Click OK to end the operation Note In Specify Dimensions from Outer Frame to Measurement Spot mm when the adja...

Page 286: ...3 The Vertical Measurement Spot Count the Horizontal Measurement Spot Count edit box become active and enter the values in it directly Maximum number of measuring spot is 999 horizontal x vertical 4 W...

Page 287: ...e from outer frame to Measurement spot and the Horizontal distance from outer frame to Measurement spot become active Enter values by using keyboard or by clicking button Valid range of vertical and h...

Page 288: ...outer frame To delete the Even split spot basing outer frame go through the following steps 1 Right click on the Standard Spot View with the Even split basing outer frame to open pop up menu And then...

Page 289: ...287 3 Specified measurement spots are deleted 4 Return step 2 and 3 to continue to delete other measurement spots...

Page 290: ...ent spots with the Even split basing outer frame To return deleted measurement pots go through the following steps 1 Right click on the Standard Spot View to open pop up menu And then select the Reset...

Page 291: ...heck box and Vertical Measurement Spot Count the Horizontal Measurement Spot Count the Vertical Dimension between Measurement Spots and the Horizontal Dimension between Measurement Spots will become a...

Page 292: ...box the setting is applied to current measured image only In the Save to All Applicable Measurement Images only the measurement image coinciding with the Trimming area of the currently displayed measu...

Page 293: ...spot file are as bellows The standard spot file can be used in SDK also 1 Open the Standard Spot Property 2 Edit the File name and Comment under the Spot file displayed is applied and click the Save b...

Page 294: ...lays Standard Spot List To display the Standard Spot List go through the following steps 1 Open the Standard Spot View 2 Right click on the Standard Spot View to open pop up menu And then select the D...

Page 295: ...rect area number Chromaticity u v correct area Judgment u v Chromaticity diagram view correct area number C h correct area Judgment L a b Color system view correct area number 4 Apply the setting to m...

Page 296: ...the measurement value on the average within the split area The following operations are performed according to the same steps Refer to the respective chapters shown below Change Display Size 5 2 2 Ch...

Page 297: ...1 Open Split Spot View To open the Split Spot View go through the following steps 1 From the Menu bar select View Split Spot View sequentially Or click the icon on the Menu bar 2 The Split Spot View i...

Page 298: ...ot To select the measurement spot go through the following steps 1 Open the Split Spot View 2 Left clicking the mouse on the view will allot the sequential numbers for the sequentially clicked lattice...

Page 299: ...View this function selects all the spots as the measurement spot To select all the measurement spots go through the following steps 1 Open the Split Spot View 2 Right click on the Split Spot View to...

Page 300: ...are sequentially allotted from the top left to bottom right according to the specified rule All cells of the data sheet are reversely displayed If the measurement spot is arbitrarily set in advance be...

Page 301: ...d or all selected measurement spots To deselect the measurement spot go through the following steps 1 Open the Split Spot View whose measurement spot is selected 2 Left clicking the mouse on the numbe...

Page 302: ...ts arbitrarily selected or all selected measurement spots To deselect the measurement spot go through the following steps 1 Open the Split Spot View 2 Right click on the Split Spot View to open Pup up...

Page 303: ...301 3 All the split spots are deselected...

Page 304: ...in a spreadsheet style To open the Split Spot Data Sheet go through the following steps 1 Open the Split Spot View 2 Right click the Split Spot View to open Pop up menu Select the item of the data to...

Page 305: ...303 3 The Split Spot Data Sheet is displayed The measurement values which saturation occur in the Tristimulus value are displayed as over and the chromaticity is displayed as error Memo...

Page 306: ...ed When you want to see data sheet of an item other than items displayed currently you have to switch the item to other see other item of the measurement data To switch the display items of the Split...

Page 307: ...l appear Select the data item to be displayed from the Switch Data Sheet Note On the Split Spot Data Sheet only one item can be displayed Multiple items cannot be displayed 4 The selected data item is...

Page 308: ...format To perform the Save CSV of the Split Spot Data Sheet go through the following steps Performing the Save CSV of the Split Spot Data Sheet saves only the currently displayed item s To save multip...

Page 309: ...elect Save CSV 4 The Explorer window will open Specify the file name with the save destination path and the file format and click Save The default file name is in the style of date time For the file f...

Page 310: ...308 5 The measurement data only displayed on the data sheet is saved Split Spot Data Sheet CSV data of Split Spot Data Sheet...

Page 311: ...he selected measurement spot into the clipboard To copy the selected measurement spot into the clipboard go through the following steps 1 Open the Split Spot View 2 Select the measurement spot on the...

Page 312: ...nu is displayed Select Copy to Clipboard The selected measurement spot is copied into the clipboard Note The measurement spot cannot be copied just by dragging on the Split Spot Data Sheet Be sure to...

Page 313: ...311 5 The selected portions on the data sheet are copied Save Split Spot Data Sheet in CSV File Format Paste Split Spot Data Sheet on Clipboard...

Page 314: ...e Split Spot Data Sheet To select the measurement spot in rectangle go through the following steps 1 Open the Split Spot Data Sheet 2 The portion not to be selected within the area selected in rectang...

Page 315: ...313 4 Right click anywhere within the Split Spot Data Sheet 5 The Pop up menu will appear Select Select Measurement Spots in Rectangle...

Page 316: ...314 6 The area selected in the rectangle is set as the measurement spot If the preset measurement spot is already included it is deselected...

Page 317: ...Spot Property Opens the Split Spot Property To open the Split Spot Property go through the following steps 1 Open the Split Spot View 2 Right click anywhere within Split Spot View to open pop up menu...

Page 318: ...ayed When the setting is completed click any button OK Enables the setting and closes this window Cancel Disables the setting and closes this window Apply Enables the setting and enables you to contin...

Page 319: ...set the split count of the split spots go through the following steps Note When the number of horizontal or vertical pixels cannot be divisible the number in edit box may be 1 In this case enter the n...

Page 320: ...rizontal and vertical pixel of trimmed image are displayed in edit box Split count number can be entered directly in edit box Max split count Vertical 144 Horizontal 256 Example When Vertical 50 pixel...

Page 321: ...y to all the loaded measurement images To save the split count changed measurement spot to all measurement images go through the following steps 1 Open Split Spot Property 2 To save the settings to al...

Page 322: ...e used To select the Split Spot File go through the following steps You can edit selected Split Spot File Click the Open to open the File Open dialog Select the Split Spot File and load it The loaded...

Page 323: ...rough the following steps Edit the File name and Comment under Spot file displayed is applied and click the Save button to save the random spot file Click the OK button to save the file in the place i...

Page 324: ...st Displays the split spot list To display the split spot list go through the following steps 1 Open the Split Spot View 2 Right click anywhere within the Split Spot View to open pop up menu Select th...

Page 325: ...andard spot Chromaticity x y correct area Judgment x y Chromaticity diagram view correct area number Chromaticity u v correct area Judgment u v Chromaticity diagram view correct area number C h correc...

Page 326: ...for the initial display and applicable measurement spot arrangement the measurement spot is set only to the center The following operations are performed according to the same steps Refer to the resp...

Page 327: ...n the Random Spot View go through the following steps 1 From the Menu bar select View Random Spot View sequentially Or click the icon on the Menu bar 2 The Random Spot View is opened When you place a...

Page 328: ...ou to freely arrange and measure up to 999 measurement spots To set circle and square measurement spots go through the following steps Measurement spot shape and size can be changed via random spot li...

Page 329: ...enu And then select Measurement Spot Input Circular Square 3 Click on a place where you want to put on a measurement spot on the Random Spot View Initial spot shape and size are specified via the Rand...

Page 330: ...Sets up to polygon measurement spot To set polygon measurement spots go through the following steps 1 Open the Random Spot View 2 Right click on the Random Spot View to open pop up menu And then sele...

Page 331: ...use at where you put on vertex of a polygon measurement spot Double click at last vertex point Up to 127 vertexes can be set Measurement spot can be laid to overlap Press Esc key in the keyboard to re...

Page 332: ...ement spot on the center of the measurement image To set the measurement spot on the center of the measurement image go through the following steps 1 Open Random Spot View 2 Right click on Random Spot...

Page 333: ...he measurement spot will be set on the center of the measurement image If any measurement spots are already set the number following the currently displayed number is allotted to the center measuremen...

Page 334: ...Moves measurement spots freely To move measurement spots go through the following steps 1 Open the Random Spot View 2 Right click on the Random Spot View to open pop up menu And then select the Measur...

Page 335: ...and click a place where you want to put on the measurement spot Press Esc key before confirming the moving to cancel the moving Coordination of measurement spot can be specified via the Random Spot Pr...

Page 336: ...t spots on the view with the same configuration To move all measurement spots with the same configuration go through the following steps 1 Open the Random Spot View 2 Right click on the Random Spot Vi...

Page 337: ...hem and click the place where you want to put on measurement spots Press Esc key before confirming the moving to cancel the moving Coordination of measurement spot can be specified via the Random Spot...

Page 338: ...elete the set measurement spot To deselect the measurement spot go through the following steps 1 Open Random Spot View 2 Right click on the Random spot view to open pop up menu And then select the Mea...

Page 339: ...t in which you want to delete on the Random Spot View Click a place outside the region When you delete middle of sequence number you can select whether you fill up a gap or not via the Random Spot Pro...

Page 340: ...t Delete selected measurement spots To delete selected measurement spot go through the following steps 1 Open the Random Spot View 2 Right click on the Random Spot View to open pop up menu And then se...

Page 341: ...39 3 Click a measurement spot to be deleted When you delete middle of sequence number you can select whether you fill up a gap or not via the Random Spot Property 5 5 11 Open Random Spot Property Memo...

Page 342: ...Spots Deselects the already set measurement spot To deselect all measurement spots go through the following steps 1 Open Random Spot View 2 Right click on Random Spot View to open pop up menu Select...

Page 343: ...341 3 Only the center measurement spot is set All the other measurement spots are deselected...

Page 344: ...the random spot measurement data in a tabular form To open the Random Spot Data Sheet go through the following steps 1 Open Random Spot View 2 Right click anywhere within Random Spot View to open pop...

Page 345: ...343 3 Random Spot Data Sheet is displayed The measurement values which saturation occur in the Tristimulus value are displayed as over and the chromaticity is displayed as error Memo...

Page 346: ...andom Spot View in which the spot pattern and spot size of the random spot are specified To open the Random Spot Property go through the following steps 1 Open Random Spot View 2 Right click anywhere...

Page 347: ...d When the setting is completed click any button OK Enables the setting and closes this window Cancel Disables the setting and closes this window Apply Enables the setting and enables you to continue...

Page 348: ...spot view The entry range is from 0 01 to 500 mm 4 To save the measurement spot definition set in the Random Spot Property to all applicable measurement images check Save to All Applicable Measuremen...

Page 349: ...e spot threshold can help extract bright area on the random spot view To set the spot threshold in the Radom spot go through the following step 1 Open the Random Spot Property 2 Check the Activate Spo...

Page 350: ...t area are masked When threshold is specified the area in which luminance is below the threshold are masked 3 When you cancel the spot threshold remove the Activate Spot Threshold check box in the Ran...

Page 351: ...ing to all measured image go through following step Open the Random Spot Property Check the Save to All Applicable Measurement Image check box and then click the Apply When the check box is OFF a sett...

Page 352: ...r not after you delete measurement spots To fill up a gap of spot number go through following step 1 Open the Random Spot Property 2 Check the Don t Fill up a gap of Spot number check box so that the...

Page 353: ...and 2 and 4 are not used even after deleting spot number 2 and 4 When Don t Fill up gap of Spot number is not checked software fill up gaps of spot number a spot number 3 turn to 2 5 turn to 3 Even w...

Page 354: ...used To select the Random Spot File go through the following steps You can edit selected Random Spot File Click the Open to open the File Open dialog Select the Random Spot File and load it The loade...

Page 355: ...steps The Random Spot File can be used in SDK Edit the File name and Comment under Spot file displayed is applied and click the Save button to save the random spot file Click the OK button to save th...

Page 356: ...t List Displays the Random spot list To display the Random spot list go through following step 1 Open the Random Spot View 2 Right click on the Random Spot View to open pop up menu And then select the...

Page 357: ...The Items for threshold is selected from X Y and Z in Tristimulus values Only one item from X Y Z can be selected Standard Spot Judgment Center difference standard spot Chromaticity x y correct area J...

Page 358: ...ist to open pop up menu When you select the Applying to Measured Image the spot settings are applied to current displayed measured image When you select the Save to All Applicable Measured Image the s...

Page 359: ...plied to all loaded measured image When you select the polygon from the circular square the first upper left vertex of a polygon is set at a center of the circular square When you select the circular...

Page 360: ...lowing operations are performed according to the same steps Refer to the respective chapters shown below Change Display Size 5 2 2 Change Display Size Change Display Color 5 2 3 Change Display Color C...

Page 361: ...359 2 The Contour View is opened While a mouse cursor is placed on the view the Tooltip is available for display Memo...

Page 362: ...Opens the property window to change the number of the contour lines To open the Contour Property go through the following steps 1 Open Contour View 2 Right click anywhere within Contour View to open...

Page 363: ...361 3 The Contour Property is displayed...

Page 364: ...362 5 6 3 Change Number of Contour lines Changes the number of contour lines To change the number of contours go through the following steps 1 Display Contour Property...

Page 365: ...the number of contour lines by pressing the spin buttons The default number of contour lines are 10 and the setting range is from 2 to 20 The contour lines equally divide the max to minimum luminance...

Page 366: ...ary position in real time The following operations are performed according to the same steps Refer to the respective chapters shown below Change Display Color 5 2 3 Change Display Color Change Items o...

Page 367: ...ted on the central intersection point of the measurement image When you trim the measurement image the definition of the cross section vies is reset In the Cruciform Cross Section form the cross secti...

Page 368: ...Cruciform Cross Section Form To display the Cross Section View go through following steps 1 Open the Cross Section View 2 Right click on the Cross Section View Pop up menu will appear and select the C...

Page 369: ...ross section lines When pressing this button checks are entered on all check boxs All delete All cross section lines are deleted without current editing cross section line All reset All cross section...

Page 370: ...specified by edit number to an arbitrary point in a view and horizontal and vertical cross section graph of Tristimulus value on that point will are displayed The cross section line keep in the same...

Page 371: ...line go through following steps 1 Open the Cross Section View 2 Right click on the Cross Section View and pop up menu will appear Select the Cross Section Setting in the pop up menu and the Cross Sec...

Page 372: ...yed Moving the mouse to any position while dragging on the view allows you to check the cross section data of the position as the graph in real time The last cross section line plotted with the mouse...

Page 373: ...romaticity diagram enabling you to visually confirm the color distribution It also enables you to zoom in the concentrated plotted zone which helps you to know the distribution more accurately The fol...

Page 374: ...372 2 The Chromaticity Diagram View is opened...

Page 375: ...go through the following steps 1 Open the Chromaticity Diagram View 2 Right click on Chromaticity Diagram View The Pop up menu will open 3 Selecting the Switch Chromaticity Diagrams from the Pop up me...

Page 376: ...76 Chromaticity Diagram displays the following view The selected chromaticity diagram is permanently retained and if the software is restarted the measurement image is displayed on the previously sele...

Page 377: ...go through the following steps 1 Open the Chromaticity Diagram View 2 Right click on Chromaticity Diagram View The Pop up menu will open 3 Selecting the Plot from the Pop up menu displays the plot typ...

Page 378: ...dom Spot View but requires you to set the measurement spot on each view If the measurement spot is not set no image is plotted on the Chromaticity Diagram View 4 For example when the Standard Spot is...

Page 379: ...20 nm pitch 1 Open the Chromaticity Diagram View 2 Right click on Chromaticity Diagram View The Pop up menu will open 3 The Pop up menu will open Select 10nm Pitch or 20nm Pitch from the Spectrum Dis...

Page 380: ...378 4 Selecting 10nm Pitch displays the following view...

Page 381: ...y area within the chromaticity diagram go through the following steps 1 Open the Chromaticity Diagram View 2 Determine the area to be zoomed in within the chromaticity diagram After clicking the start...

Page 382: ...Chromaticity Diagram View 4 The Pop up menu will open Select the Zoom ON 5 The zoomed in area within the view is enlarged Selecting the property of the Pop up menu on the zoomed in display allows you...

Page 383: ...he operation of zooming in the chromaticity diagram To cancel the operation go through the following steps 1 Open the zoomed in Chromaticity Diagram View 2 Right click anywhere within Chromaticity Dia...

Page 384: ...382 4 The view returns to the original display...

Page 385: ...Chromaticity Diagram and CIE1976 Chromaticity Diagram on the Chromaticity Diagram View To turn ON the color display go through the following steps 1 Open the Chromaticity Diagram View 2 Right click a...

Page 386: ...384 4 When Color ON is selected the Chromaticity diagram is colored...

Page 387: ...lor of the CIE1931 Chromaticity Diagram and CIE1976 Chromaticity Diagram To turn OFF the color display go through the following steps 1 Open the Chromaticity Diagram View 2 Right click anywhere within...

Page 388: ...386 4 When Color OFF is selected the Chromaticity diagram is display with monotone...

Page 389: ...r judgment result by setting coordinate values To set judgment area go through the following steps 1 Open the Chromaticity Diagram View 2 Right Click on Chromaticity Diagram View The pop up menu wll o...

Page 390: ...t Click selecting Judgment Area Pop up menu will open Select Add to add judgment aera to list on right 5 Select circle or rectangle for the shape and input coordinate values by key After that click Ap...

Page 391: ...389 By adding judement area and or child area cotinuouly overlayed areas will be built Showing judgment area...

Page 392: ...ity xy and or u v area to be applied for judgment result by mouse To set judgment area go through the following steps 1 Open the Chromaticity Diagram View 2 Please right click and Pop up menu will ope...

Page 393: ...391 3 Please draw shapes by dragging mouse and click mouse again to finish editing 4 Information about about will be influenced on Choromaticity Judgment Area dialog...

Page 394: ...392 By drawing area and or child area cotinuouly overlayed areas will be built...

Page 395: ...xy and or u v judgment area To remove judgment area go through the following steps 1 Open Chromaticity Diagram View 2 Please right click and Pop up menu will open 3 Please select Judgment Area Setting...

Page 396: ...lowing steps 1 Open Chromaticity Diagram View that has judgment area 2 Please right click on Chromaticity Diagram View Pop up menu will open 3 By selecting Remove Judgment Area and click shape in the...

Page 397: ...nd or u v judgment area To move judgment area go through the following steps 1 Open Chromaticity Diagram View that has judgment area 2 Please right click on Chromaticity Diagram View Pop up menu will...

Page 398: ...396 4 Judgment area will turn active and movable By clicking destination in the view you can finish moving...

Page 399: ...e for chromaticity xy and or u v judgment To select file for judgment area go through the following steps 1 Please show Chromaticity Diagram View Property 2 By clicking Open button dialog will open Pl...

Page 400: ...t area go through the following steps 1 Please show Chromaticity Diagram View Property 2 Current file is shown under the text of File currently being displayed is applied Please edit File Name and Com...

Page 401: ...romaticity Diagram Coordinates Use this function to zoom in the arbitrary position of the chromaticity diagram or display the specific area To open the Chromaticity Diagram View Property go through th...

Page 402: ...400 4 The Chromaticity Diagram View Property is displayed...

Page 403: ...his function to zoom in the arbitrary position of the chromaticity diagram or display the specific area If the Zoom function is enabled the zoomed in area range is displayed as the default The procedu...

Page 404: ...ue and interval of the longitudinal axis and lateral axis the data for the zoomed in range are used as the initial values 5 Change the minimum value or maximum value to zoom in or out the current view...

Page 405: ...n the set maximum and minimum values 7 To change the interval of the axis auxiliary line enter the interval To activate the change press the OK button Entry range of maximum value and minimum value fo...

Page 406: ...ls Longitudinal axis For the CIE1931 Chromaticity Diagram enter chromaticity y value For the CIE1976 Chromaticity Diagram enter chromaticity v value Lateral axis For the CIE1931 Chromaticity Diagram e...

Page 407: ...iew function enables you to visually understand the Tristimulus value distribution The following operations are performed according to the same steps Refer to the chapter shown below Save Snapshot 5 2...

Page 408: ...406 2 The Histogram View is opened...

Page 409: ...s you to confirm the Tristimulus value distribution three dimensionally The following operations are performed according to the same steps Refer to the chapter shown below Save Snapshot 5 2 11 Save Sn...

Page 410: ...408 2 The 3D View is opened Dragging on the view enables you to freely handle the 3D view at any angle and operating the mouse wheel allows you to zoom in or out the view...

Page 411: ...l measurement images Selecting the downsized measurement image on the view enables you to switch the displayed measurement image In the Thumbnail View the images are displayed sequentially arranged fr...

Page 412: ...410 2 The Thumbnail View is opened...

Page 413: ...humbnail View represents the currently displayed measurement image Left click the mouse to select the measurement image It is also possible for you to move the cursor using the keyboard arrow keys and...

Page 414: ...or View It displays the measurement data in RGB color that is close to color of measuremet object itself To open RGB View go through the following steps 1 From the menu bar select the View RGB Color V...

Page 415: ...RGB Color View Property will open Set color tones Set the gradation according to the specified maximum luminance When you want to terminate the setting operation click one of the following buttons OK...

Page 416: ...x for To specify the maximum Luminance 2 Enter the maximum luminance value If the maximum luminance value entered becomes red it can not display RGB Color View because it is an invalid value When the...

Page 417: ...a b View Displays L a b View It displays measuremet data in CIE L a b values To open L a b View go through the following steps 1 From the menu bar select the View L a b View sequentially or clicke the...

Page 418: ...rformed according to the same steps Refer to the respective chapters indicated below Standard Spot View Operation 5 3 Standard Spot View Operation Split Spot View Operation 5 4 Split Spot View Operati...

Page 419: ...e of L a b View To change scale go throught the following steps 1 From the menu bar select the View L a b View sequentially 2 Please right click in the L a b View pop up menu will open Select menu lik...

Page 420: ...418 3 Scale Setting dialog will open You can change display scale of L a b view...

Page 421: ...splay color of L a b View To switch color go through the following steps 1 From the menu bar select the View L a b View sequentially 2 Please right click in the L a b View pop up menu will open Select...

Page 422: ...or pass or fail judgment To set judgment area go through the following steps 1 Open L a b View 2 Please right click in the L a b View pop up menu will open 3 Please select Judgment Area Setting from p...

Page 423: ...421 4 Right Click selecting Judgment Area Pop up menu will open Select Add to add judgment aera 5 Input values for h1 h2 C 1 C 2 by key Click OK or Apply to finish complete changes...

Page 424: ...422 h1 Lower Hue Area deg 0 360 h2 Upper Hue Area deg 0 360 C 1 Lower Chroma Area 0 150 C 2 Upper Chroma Area 0 150...

Page 425: ...a b judgment area To remove L a b judgment area go through the following steps 1 Open L a b View 2 Right click in L a b View Pop up menu will open 3 After selecting Judgment Area Setting from pop up m...

Page 426: ...king Reset judgment area in selected area number will be removed By clicking Reset All all of judgment areas will be removed Judgment areas can also be canceled by using Delete of the pop up menu in t...

Page 427: ...splays measurement data of L a b View in spread sheet To open L a b Data Sheet go through following steps 1 Open L a b View 2 Right click in the L a b View Pop up menu will open Select Display Data Sh...

Page 428: ...isplay Items To switch display item of L a b Data Sheet go through the following steps 1 Open L a b Data Sheet 2 Right click in the L a b Data Sheet 3 Pop up menu will open Select data item to be disp...

Page 429: ...5 14 1 Open Hue Chroma Color system View Displays Hue Chroma View To open Hue Chroma View go through the following steps 1 From the menu bar select the View Hue Chroma View sequentially or click the i...

Page 430: ...ot are performed according to the same steps Refer to the respective chapters indicated below Standard Spot View Operation 5 3 Standard Spot View Operation Split Spot View Operation 5 4 Split Spot Vie...

Page 431: ...Changes display scale of Hue Chroma View To change scale go through following steps 1 Open Hue Chroma View 2 Please right click in the Hue Chroma View pop up menu will open Select menu like Scale Sett...

Page 432: ...430 3 Scale Setting dialog will open You can set scale values by key...

Page 433: ...und color Changes background color of Hue Chroma View To change background color go through the following steps 1 Open Hue Chroma View 2 Right click in the Hue Chroma View Pop up menu will open Select...

Page 434: ...nt of Hue Chroma View in spread sheet To open Hue Chroma Data Sheet go through following steps 1 Open Hue Chroma Data Sheet 2 Right click in the Hue Chroma Data Sheet Pop up menu will open Select Disp...

Page 435: ...o Color View Displays Color System Pseudo Color View To open Color System Pseudo Color View go through the following steps 1 From the menu bar select the View Color System Pseudo Color View sequential...

Page 436: ...e display mode go through the following steps 1 Right click in the Color System Pseudo Color View Pop up menu will open Select menu item like Displaying Color System L 2 Display mode will switch to th...

Page 437: ...peration 5 16 1 Open Judgment result View Displays Judgment Result View To open Judgment Result View go through the following steps 1 From the menu bar select the View Hue Chroma View sequentially or...

Page 438: ...436 2 Judgment Result View will open If Judgment was acctepted spots will be colored in green If rejected spots is colored in red...

Page 439: ...are performed according to the same steps Refer to the respective chapters indicated below Standard Spot View Operation 5 3 Standard Spot View Operation Split Spot View Operation 5 4 Split Spot View O...

Page 440: ...2 Right click in the Judgment Result View Pop up menu will open Select Judgment Result List 3 Judgment Result List will open Accepted spots in judgment are colored in green Rejected spotw are colored...

Page 441: ...nt conditions Sets judgment condition To set judgment condition go through the following steps 1 Open Judgment Result View 2 Right click in the Judgment Result View Pop up menu will open Select Judgme...

Page 442: ...it Lower Limit Make a judgment by upper and lower value Value of spot should be within these values Ratio to max Make a judgment by ratio to maximum value in spot Center diffence Make a judgment by di...

Page 443: ...ejected judgment To select a setting file for judgment go through the following steps 1 Open Judgment Result View and right click Pop up menu will open Select Judgment condition setting 2 Click Open t...

Page 444: ...Judgment Result View and right click Pop up menu will open Select Judgment condition setting 2 Current file is shown under the text of File currently being displayed is applied Please edit File Name a...

Page 445: ...go through the following steps For the measurement spot type of the Time series Measurement View select either Standard Spot Split Spot or Random Spot For the measurement spot setting of each type re...

Page 446: ...surement View to open Pop up menu Select one from Random Spot Split Spot and Standard Spot via the Switch Time series Displays About the operation of Random Spot Split Spot and Standard Spot refer to...

Page 447: ...splayed This function is used to open the time series graph again after closing it To open the time series graph go through the following steps Open Time series Layout 6 4 2 Open Time series Layout 1...

Page 448: ...446 3 The Time series Graph is displayed...

Page 449: ...asurement spot list go through the following steps The operational procedures are the same in all Standard Spot Split Spot and Random Spot of the Time series Measurement View 1 Display the Time series...

Page 450: ...Split Spot List You can directly edit Comment for the measurement spot set The layout of the Time Series Measurement Spot List are different from Spot view Time series Measurement Standard Spot List T...

Page 451: ...449 Time series Measurement Random Spot List...

Page 452: ...t displayed on the time series measurement graph To select the spot list go through the following steps 1 Open the Time series Measurement Spot List The items with the number checkbox checked in the m...

Page 453: ...ement spot list If 10 measurement spots are already selected the measurement spot cannot be additionally selected After deselecting the measurement spot not to be displayed check the checkbox of the s...

Page 454: ...urity are displayed on the left and right longitudinal axes The left longitudinal axis item is graph displayed as a solid line while the right longitudinal axis item is graph displayed as a dashed lin...

Page 455: ...ry area within the graph go through the following steps 1 Activate the Time series Graph 2 Determine the area to be zoomed in within the Time series Graph After clicking the starting point drag the mo...

Page 456: ...454 4 The Pop up menu will open Select the Zoom ON 5 The specified area is enlarged...

Page 457: ...of the arbitrary area of the Time series Measurement Graph To cancel the zooming in of the arbitrary area within the graph go through the following steps 1 Open the zoomed in Time series Graph 2 Righ...

Page 458: ...456 4 The zoomed in area returns to the original display...

Page 459: ...n order to change the axis scale for the time series measurement graph To open the time series measurement graph property go through the following steps 1 Open the Time series Graph 2 Right click anyw...

Page 460: ...is displayed When the setting is completed click any button OK Enables the setting and closes this window Cancel Disables the setting and closes this window Apply Enables the setting and enables you t...

Page 461: ...zoom operation is executed 1 Open the Time series Measurement Graph and Time series Measurement Graph Property The example is based on the case when using the Zoom function on the Time series Measure...

Page 462: ...alue Maximum value 1 to 9999999 Minimum value 0 to 9999998 Chromaticity Maximum value 0 0001 to 0 999999 Minimum value 0 0000 to 0 999998 Color temperature Maximum value 1564 to 100000 Minimum value 1...

Page 463: ...Since the entry range is not specified the values can be entered even if they are out of the graph plot Therefore if the entry range is entered incorrectly the plot is not displayed Selecting the Aut...

Page 464: ...462 When Apply is pressed after changing the property scale the scale range is changed...

Page 465: ...ances in a spreadsheet style 5 19 1 Switch Display Items of Data Sheet On the Time series Measurement Data Sheet one displayed item can be selected To switch the displayed items go through the followi...

Page 466: ...menu and the display item list are displayed Select the display item Only one item can be displayed When the item is set the setting is applied to Time series Measurement Graph to display the item on...

Page 467: ...V File Format Saves the contents of the data sheet in CSV or TEXT file format To save the data sheet go through the following steps 1 Open the Time series Measurement Data Sheet 2 Right click anywhere...

Page 468: ...4 The Explorer window will open Specify the path file name and file format for the file to be saved The default file name is in the style of date time For the file format you can select the CSV or TE...

Page 469: ...467 5 The following is the status when a window is opened by Excel The content displayed on the data sheet is saved Time series Measurement Data Sheet...

Page 470: ...pen from File Menu Opens the measurement image file from the File menu To open the measurement image from the File menu go through the following steps 1 From the Menu bar select File Open Measurement...

Page 471: ...open Increasing the number of measurement images will take more time 4 Measurement Image Load List dialog will open The measurement date time and comment of the loaded measurement image are displayed...

Page 472: ...number limited to 999 including the number of currently retained files However if a large number of measurement image files are opened at once it may take more time to completely open the files If yo...

Page 473: ...10 software 2 Open the folder where the measurement image is saved After opening the folder select the measurement image to be loaded and drag and drop the image on the main window When opening the me...

Page 474: ...e and comment of the loaded measurement image are displayed To open the file click OK 5 The loaded measurement image is displayed Double clicking the measurement image file also enables you to open th...

Page 475: ...ent image to be saved such as a trimmed measurement image the following dialog will appear To save the file select Cancel to save the measurement image Open Time series Layout 6 3 Save Measurement Ima...

Page 476: ...dition The measurement image which was not yet saved after the measurement When the saved measurement image is trimmed When the standard spot of the saved measurement image is changed When the random...

Page 477: ...image To save the measurement image go through the following steps 1 From the Menu bar select File Save Measurement Image 2 The Enter Comment dialog is displayed If you want enter the comment And the...

Page 478: ...reate a new folder click Create New Folder A new folder is created Enter the folder s name Specify the folder containing the measurement image s Folders on the network can be selected 4 If it will tak...

Page 479: ...aves the measurement images To save the measurement images go through the following steps 1 From the Menu bar select File Save All Measurement Images 2 Measurement Image List dialog will appear Edit a...

Page 480: ...the Measurement Image List and right click them and select the Comment Editing Comment entered in the Enter Comment dialog is entered to all selected columns When resaving the image file the previous...

Page 481: ...be entered comment in the Measurement Image List and right click them and select the Comment Editing Select the relevant image on the Measurement Image List and right click to select Edit File Name Th...

Page 482: ...create a new folder click Create New Folder A new folder is created Enter the folder s name Specify the folder containing the measurement image s Folders on the network can be selected 6 If it will t...

Page 483: ...eps 1 From the Menu bar select File Save Image As 2 The Enter Comment dialog is displayed If you want enter the comment And then click OK Leave the comment field blank and click OK if you don t need t...

Page 484: ...ay the initial layout go through the following steps 1 From the Menu bar select View Initial Layout sequentially 2 The Initial Layout window is displayed If you select the menu while the current measu...

Page 485: ...t image is displayed all the windows will close and the measurement image will be redisplayed in the Time series Layout Time series layout is a fixed layout for Time series Measurement View Time serie...

Page 486: ...nt field the comment set through Save Measurement Image or Save All Measurement Images is displayed To open the measurement image list go through the following steps 1 From the Menu bar select View Me...

Page 487: ...File Name Edits the file name of the measurement image To edit the file name of the measurement image go through the following steps Select the measurement data to be edited and select Edit File Name...

Page 488: ...486 6 5 2 Comment Editing Edit a comment for measurement image To edit comments for each measurement data go through following steps 1 Click a comment column and enter comment...

Page 489: ...same comment in several column select columns to be entered comment in the Measurement image List and right click them and select the Comment Editing Comment entered in the Enter Comment dialogue is...

Page 490: ...urement Data Delete Deleting measurement data in memory To delete measurement data in memory go through following steps Select measurement data to be deleted and select the Measurement Data Delete wit...

Page 491: ...Image Displaying a measurement recipe To display a measurement recipe go through following steps Select a measurement recipe to be displayed and then select the Recipe of Measurement Image with right...

Page 492: ...rement image on the display To change measurement image go through following steps Select a measurement image to be displayed and then select the Switch the Displaying Measurement Image with right cli...

Page 493: ...ck the measurement conditions To open the recipe of the currently displayed measurement image go through the following steps 1 From the Menu bar select View Recipe of Currently Displayed Measurement I...

Page 494: ...top left top right bottom left and bottom right This menu is intended just for classifying arbitrary windows to be displayed and is functionally different from Initial Layout To organize the scattered...

Page 495: ...493 2 From the Menu bar select Window Tile sequentially 3 The displayed views are organized...

Page 496: ...The measurement image is kept in the loaded condition and is not deleted This menu is intended just for closing the window and is functionally different from Close Measurement Image 1 The following sc...

Page 497: ...495 2 From the Menu bar select Window Close All sequentially...

Page 498: ...plays the Help for all the functions 8 1 1 Open the Instruction Manual from HELP Menu To use the Help on the Menu bar go through the following steps 1 From the Menu bar click Help Topic Search sequent...

Page 499: ...you contact Topcon when you have any problem in the software To open the version information go through the following steps 1 From the Menu bar click Help Topic Search sequentially 2 The About dialog...

Page 500: ...498 View mode UA 10 Series UA 200 Series...

Page 501: ...whether the computer operating environment is adequate If other software is operated be sure to exit the software and then restart this software Application is already running Details This software is...

Page 502: ...ice When RS 232C is utilized as the communication method there may be something wrong in the communication setting and port number Countermeasure Check for the following matters The device driver is i...

Page 503: ...specification and then re execute the measurement Failed to calculate an optimal value because the light source of the measurement object was too weak Details Even when the integral time is set at th...

Page 504: ...s The DPI is not set to 96 Countermeasure Select the Display Property Settings tab Advanced Settings General tab in OS configuration sequentially set the DPI Setting on the General tab to 96 DPI and t...

Page 505: ...oes not satisfy with recommended operating condition the communication error may occur in PC or device In this case change the setting of packet size and packet delay Packet size Packet delay and memo...

Page 506: ...ftware The procedure for starting up GigEConf exe is as follows 1 Connect PC to the device and then turn on the device 2 Select the GigEConf via Start All Program TOPCON TECHNOHOUSE UA 10_200 Then Gig...

Page 507: ...505 UA 200A...

Page 508: ...ries only 1 Select a device information displayed on the dialog 2 Selected device s IP address Subnet mask and Default gateway are displayed on the each edit box and combo box When one or more devices...

Page 509: ...tting is completed following dialog will appear When setting is not completed following dialog will not appear 5 Turn off and on the power of the device Setting take effect after power off and on of t...

Page 510: ...g are available for UA 10 series only 1 Select a device information displayed on the dialog 2 Selected device s Packet size Packet delay and Memory channel are displayed on the each edit box and combo...

Page 511: ...ot completed following dialog will not appear Setting takes effect immediately so there is no need to turn off and on Memory channel 1 is manufacturer s default When you want to set it to manufacturer...

Page 512: ...he values in practice The measurement distance is from tip of objective lens to the measurement target Above mentioned measurement area is 90 of field of view Luminance measurement range 0 1 30000cd m...

Page 513: ...Standard illuminant A evaluated on the center of the CCD Repeatability Luminance 2 Standard illuminant A evaluated on the center of the CCD Temperature characteristic Luminance 3 0 to 40o C reference...

Page 514: ...t UA 10SL 90 0 Length 38 5 Width 38 5 Height UA 10SH 85 1 Length 34 0 Width 34 4 Height UA 10T Weight About 0 170kg UA 10W About 0 185kg UA 10SL About 0 200kg UA 10SH About 0 155kg UA 10T The measurem...

Page 515: ...istance is from tip of objective lens to the measurement target Above mentioned measurement area is 90 of field of view Luminance measurement range 0 05 1000000cd m2 Luminance linearity 2 Standard ill...

Page 516: ...ticity including transfer time Stability Luminance 1 Standard illuminant A evaluated on the center of the CCD Repeatability Luminance 2 Standard illuminant A evaluated on the center of the CCD Tempera...

Page 517: ...th 162 0 Width 194 0 Height UA 200WS 307 1 Length 162 0 Width 194 0 Height UA 200T Weight About 3 90kg UA 200S About 4 12kg UA 200WS About 4 45kg UA 200T The measurement condition of each performance...

Page 518: ...on the center of CCD Chromaticity accuracy 0 003 Standard illuminant A 1cd m2 evaluated on the center of CCD 0 005 Standard illuminant A 1cd m2 evaluated on the center of CCD 0 008 Standard illuminan...

Page 519: ...100cd m2 Luminance and Chromaticity including transfer time Stability Luminance 1 Standard illuminant A evaluated on the center of the CCD Repeatability Luminance 2 Standard illuminant A evaluated on...

Page 520: ...162 0 Width 194 0 Height UA 200AWS 307 1 Length 162 0 Width 194 0 Height UA 200AT Weight About 3 90kg UA 200AS About 4 12kg UA 200AWS About 4 45kg UA 200AT The measurement condition of each performan...

Page 521: ...mply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is ope...

Page 522: ...520 External Dimension Diagram UA 10 series Main body with Tripod screw adapter...

Page 523: ...521 Depth5 Main body with Tripod screw adapter and Spacer Mount...

Page 524: ...522 UA 200 series UA 200S UA 200AS...

Page 525: ...523 UA 200WS UA 200AWS...

Page 526: ...524 UA 200T UA 200AT...

Page 527: ...er date in the data analysis etc The measurement image files with the extension msr cannot be opened by any software other than this software Opening the file forcedly may cause it to be damaged so ne...

Page 528: ...e exceeds the measurable upper limit of the photo detector CCD Even if the brightness exceeding this upper limit is measured the CCD output value remains at the upper limit and therefore the data may...

Page 529: ...onditions make the time longer ND Filter Neutral density filter Optical filter to reduce the amount of light incoming CCD This filter protects happening saturation when UA 200 and UA 200A measure very...

Page 530: ...at the pixel Chromaticity x y u v Chromaticity at the pixel If saturation occurs over is displayed Pop up menu The menu appear after right click each views and sheets The menu vary from each views and...

Page 531: ...he target pixel value 1000 are obtained When these values are arranged from the lowest in order they are arranged as shown below 20 30 40 50 60 70 80 90 1000 The middle value is 60 The middle value 60...

Page 532: ...aussian filter which is determined by taking the distance and difference in values from each pixel into consideration is applied This method is effective to remove noise without blurring profile edge...

Page 533: ...rable Do not hesitate to call our local retailer where you purchased the product or TOPCON TECHNOHOUSE CORPORATION 1 The repair parts are the parts necessary to keep the product functions 2 Although w...

Page 534: ...vice section Phone 81 3 3558 2710 Fax 81 3 3558 3011 Luminance and Chromaticity Uniformity Analyzer UA 10 series 2D Luminance Colorimeter UA 200 series Instruction Manual Publication date November 201...