36
(2)Adjusting Calibration
Turn the unit off by pressing the [START] pad once. Confirm that unit has shut off before
beginning the following procedure.(In step 2 and 3, optional RC-30 remote control is helpful.)
1
Without moving the unit, press the [ENTER] and [START] keys simultaneously.
2
Press either the right or left key on the Alignment Control pad to move the laser beam up
or down on wall B until the measurement for the distance dHb is the same as the mea-
surement dHa on wall A.
• This operation can be performed with the Remote Controller RC-30.
When using the RC-30, select the Y axis using the X/Y axis selection key and
move the laser beam up and down with the Alignment control key.
START
1
X
V
Y
2
3
ENTER
ERROR
4
X / Y
Laser beam
up or down