32
3
When the beam is precisely centered, press the [ENTER] key.
The [X] LED will flash.
When the flashing stops, the X axis calibration adjustment is made and power is turned off.
4
For Y axis calibration, turn the unit as instructed in step 1 then press the X/Y Axis Selection
pad.
Confirm that the [Y] LED is lit, then repeat steps 2 to 4 for the Y axis.
Repeat the checking procedure to confirm proper calibration has been made.
• This operation can be performed with the Remote Controller RC-30.
When using the RC-30, press the [ENTER] key on the RC-30 unit.
(The [ENTER] key for the RC-30: The same mark as the [ENTER] key for the in-
strument.)
• This operation can be performed with the Remote Controller RC-30.
When using the RC-30, select the Y axis using the X/Y axis selection key and
move the laser beam up and down with the Alignment control key.
Note
• If the calibration is greater than the adjustment allows, the error LED will start
flashing. (See “Error Code” on
If this occurs, contact your Topcon dealer.
• When the laser beam is difficult to see, it can be adjusted in the Laser sensor
mode or the Laser pointing mode.