15. RESECTION MEASUREMENT
70
4. Sight the first known point and press
[MEAS]
to
begin measurement.
The measurement results are displayed on the
screen.
• When
[ANGLE]
has been selected, the distance
cannot be displayed.
5. Press
[YES]
to use the measurement results of
the first known point.
• You can also input target height here.
• Press
[NO]
to return to the screen in step 4 and
perform measurement again.
6. Repeat procedures 4 to 5 in the same way from
subsequent points.
When the minimum quantity of observation data
required for the calculation is present,
[CALC]
is
displayed.
7. Press
[CALC]
or
[YES]
to automatically start
calculations after observations of all known points
are completed.
• Instrument station coordinates, station elevation,
and standard deviation, which describes the
measurement accuracy, are displayed.