20. CHANGING THE SETTINGS
• The setting for "Fine AVG" distance measurement mode can be increased/decreased using the
[ + ]
/
[ - ]
softkeys.
• Target information can be edited and recorded.
"PROCEDURE Recording and editing target information"
• "Prism constant" will not be displayed when "N-Prism" is selected in "Reflector".
• When "prism constant" values are changed and
[OK]
is pressed, these changes are reflected in
the status bar/Starkey mode target type display temporarily. This display will also change to
reflect changes in target information settings made using a data collector temporarily. In both of
the above cases, changes will not be recorded in <Reflector setting>.
Status bar: "5.2 Display Functions", Starkey mode: "5.4 Starkey Mode", <Reflector setting>:
"PROCEDURE Recording and editing target information", Cold boot: "9.2 Resolving
Software Issues
• The Guide light brightness item ("Guide light (Brightness)") will be displayed only when "Illum.
hold" is set to "Guide light".
Prism constant correction
Reflective prisms each have their prism constant.
Set the prism constant correction value of the reflective prism you are using. When selecting "N-
Prism" in "Reflector", prism constant correction value is set to "0" automatically.
"ppm" tab
•
[0ppm]
: Atmospheric correction factor returns to 0 and temperature and air pressure are set to
the factory settings.
• Atmospheric correction factor is calculated and set using the entered values of the temperature
and air pressure. Atmospheric correction factor can also be entered directly.
• When the settings here are different between basic mode and program mode, program mode
settings are prior for use.
Items set, options, and input range (*: Factory setting)
Temperature:
-30 to 60°C (15*)/-22 to +140°F (59*)
Pressure:
500 to 1400hPa (1013*)/375 to 1050mmHg (760*)
/
14.8 to 41.3 (inchHg) (29.9*)
Humidity:
0 to 100% (50*)
ppm (Atmospheric correction factor):
-499 to 499 (0*)
•The "Humidity" item is displayed only when the "ppm setting" in "Obs. condition" is set to
"+Humidity".