CA-200F Corneal Analyser - Rev. 7 16/01/2012
62
Right alignment
Wrong alignment
fig. 6-2
From the CA-200F software, select the
"Settings"
button and then the
“Calibration Check"
function (
fig. 6-3
Following the instructions given at the bottom of the window, acquire the sphere eight times with a radius of
8 mm.
If the calibration is correct, a window will appear with the message "Calibration Check: POSITIVE".
If the test has a negative outcome, retry. If the outcome is negative for three consecutive times, contact
Technical Service to have the device recalibrated.
To interrupt calibration and return to the acquisition environment, press the
“Close”
button.