41
2.7.2 Analog Output Response Speed
The analog output response speed differs depending on the range used.
Response Speed
☞
「
Appendix: Response Speed
」
When checking the build up characteristics when a light is on, set the FAST/SLOW switch to
FAST. (In the case of SLOW, the integral capacitor value is increased and operation is
performed smoothly, so the response speed becomes slow.)
2.7.3 Usage
The procedure for analog output measurement is as described below.
1
Solder and attach the shielded wire of the supplied analog output plug, and connect it to
the device.
・
The pin side of the plug is the signal.
2
In the function mode, set the respective measurement ranges for X, Y, Z.
Range Setting
☞
「
3.8 Measurement Range
」
3
Press the panel switch AUTO/MANU, and make sure that [MANU] of the panel switch
blinks. The manual range state is set
4
Set FAST/SLOW depending on the measurement purpose.
FAST/SLOW Setting
☞
「
1.2 Name and Function of Each Part
」
Request
・
When an absolute value is necessary for the tristimulus value, take the
correlation with the display into consideration.
・
When used with the output voltage in the saturated state, proper
measurement cannot be performed. Be sure to set the proper
measurement range.
・
The output impedance is about 100
Ω
. Use a recorder having an input
impedance of 10k
Ω
or more.
・
The zero point may move due to temperature change, but this does not means the
instrument is not broken.