3.8 Measurement Range
When using the manual range, the measurement range for X, Y, and Z is displayed.
Display Method
☞
「
3.1.2 Data/Setting Item Display
」
The procedure for making changes is as described below.
1
Press the [CHANGE] button.
It is no possible to select the value for X.
2
Press the [ROTATION] button and select the setting contents
3
After setting is completed, press the [ENTER] button.
Now, the value for Y can be selected, so similarly select the value for Y and Z.
・
The measurement range is from 1 to 5, with range 1 having the maximum sensitivity.
Measurement Range Details
☞
「
Appendix: Specifications
」
・
When the AUTO/MANU switch is set to AUTO, and dip switch 1 is set to the remote side,
the measurement range setting is invalid.
・
The factory settings for X, Y and Z are all range 3.
・
After selecting Z, press the [ENTER] button. The next setting item (screen) is displayed.
*
Measuring
range
*
X
:
Range1
Y
:
Range3
Z
:
Range3