7
1.2.4.1 Maximum and minimum values for different ranges
Frequency
Parameter
100Hz
120Hz
1kHz
10kHz
Cmax
800
μ
F
667
μ
F
80
μ
F
8
μ
F
Cmin
1500pF
1250pF
150pF
15pF
Lmax
1590H
1325H
159H
15.9H
Lmin
3.2mH
2.6mH
0.32mH
0.032mH
Zmax/
Rmax
1M
Ω
Zmin/ Rmin
1.59
Ω
Table 1-2. Maximum and Minimum Values
1.2.4.2 Deviation factor ks
slow speed: ks =0
fast speed: ks =10
1.2.4.3 Level factor kv
When V=1V, kv =0;
When V = 0.3V, kv = 1;
When V = 0.1V, kv = 4.
1.2.4.4 Frequency factor kf
When f = 100 Hz, 120 Hz or 1 kHz, kf = 0;
When f = 10 kHz, kf = 0.5.
1.2.5 Test signal frequency
Test signal: sine wave
Four frequencies are available: 100Hz, 120 Hz, 1 kHz, 10 kHz
Accuracy: 0.02%.
1.2.6 Test signal level
Three levels are available: 0.1V, 0.3V, 1.0V (effective value)
Accuracy: ±10% (The set value of the measurement level is the value set when
test terminals are opened HD and HS should be connected)
Due to the output impedance of the signal source, the actual level applied upon
the component under test may be different from the set value.
1.2.7 Output impedance of measurement signal
Output impedance: 30
Ω
(default value) and 100
Ω
, without requirement of
accuracy.