8
1.2.8 Measurement speed
Fast: around 15 times/sec.;
Slow: around 4.5 times/sec.;
The actual speed differs with different measurement conditions such as
frequency, component value, display mode, measurement mode, once, range
mode, RS232 interface and so on.
The typical speeds for Fast, Medium and Slow are given under the following
measurement conditions:
·
Frequency: 1 kHz or 10 kHz;
·
Measurement mode: continuous;
·
Range mode: Hold;
·
Display parameters: C/D;
·
RS232 interface: only receiving allowed.
1.2.9 Inputting data range
Parameter Code
Range
L
0.0001
μ
H
~
99999H
C
0.0001pF
~
99999
μ
F
Nominal value
R/|Z|
0.0001
Ω
~
99999M
Ω
High or low limit
-9999%
~
9999%
Dissipation value
Quality factor
D/Q
0
~
99999
Table 1-3. Sorting Setup Range
1.3 Operating
environment
1.3.1 Temperature & Humidity
Temperature: 0
0
C ~ 40
0
C
Humidity:
≤
85
%
RH
Accurate temperature: 18
0
C ~ 28
0
C
Accurate humidity:
≤
80
%
RH
1.3.2 Power
Voltage: 220V
(
1±10
%)
Frequency: 50Hz
(
1±5
%)
Power: around 25W
1.3.3 Warm-up
After warming up for 10 minutes, the instrument can start to measure.