26
suggested that limit parameter and nominal value be input according to the
format requirement after the instrument is adjusted, in order to avoid being
dealt with as invalid command.
No. Command
code
Note 0
Characters
Function
1
Note 1
A0
2
measurement parameter & series
connection and parallel
connection: L-Q
2
Note 2
A1
2
measurement parameter & series
connection and parallel
connection: C-D
3
A2
2
measurement parameter & series
connection and parallel
connection: R-Q
4
A3
2
measurement parameter & series
connection and parallel
connection: Z-Q
5
B0
2
measurement frequency: 10 kHz
6
B1
2
measurement frequency: 1 kHz
7
B2
2
measurement frequency: 120 Hz
8
B3
2
measurement frequency: 100 Hz
9
C0
2
measurement level: 1 V
10
C1
2
measurement frequency: 0.3 V
11
C2
2
measurement frequency: 0.1 V
12 D0
2
display
mode:
%
△
13
D1
2
display mode: direct
14
E0
2
range: holding at current range
15 E1
2
range:
auto
16 E2
2
range: holding at 0 (100 k
Ω
)
17 E3
2
range: holding at 1 (10 k
Ω
)
18 E4
2
range: holding at 2 (1 k
Ω
)
19 E5
2
range: holding at 3 (100
Ω
)
20 E6
2
range: holding at 4 (31.6
Ω
)
21 E7
2
range: holding at 5 (10
Ω
)
22
F0
2
measurement speed: fast
23
F1
2
measurement speed: slow
24 G0
2
short
correction
25 G1
2
open
correction
26
H0
2
alarm: (ON) to turn on beeper
27
H1
2
alarm: (OFF) to turn off beeper
28
I0
2
trigger: (Off) continuous
29
I1
2
trigger: (Off) single
30
J0
2
equivalent model: (SER) serial