27
connection
31
J1
2
equivalent model: (PAR) parallel
connection
32
K0
2
RS232A
口
: (OFF) only receiving
33
K1
2
RS232A
口
: (ON) receiving and
sending
34
L0
2
sorting mode: P1
35
L1
2
sorting mode: P3
36
M0
2
signal source resistance: 30
Ω
37
M1
2
signal source resistance: 100
Ω
NX=X
1
X
2
BX
3
X
4
X
5
X
6
X
7
≤
10
NX: related set parameters
Parameter values and units are
after “=”.
The regulations are as follows:
N1: nominal value N2: Q/D limit
value
N3: first-class high limit N4:
first-class low limit
N5: second-class high limit N6:
second-class low limit
N7: third-class high limit N8:
third-class low limit
Data format is as follows after
being received and adjusted:
1. N1: X1: ASCII code (0
∼
9);
X
2
∼
X
6
: ASCII code (0~9 and
point);
X
7
: ASCII code (0
∼
2) (unit code)
Definition of X
7
is as follows:
X
7
’s ASCII
code
C L Z,
Q
30H
pF
μ
H
Ω
31H
nF
mH
K
Ω
32H
μ
F
H
M
Ω
38
Example and notes: (symbols used to
express in the examples, and ASCII
code should be used in real-world
program)
1. N1=122 will be automatically
adjusted to N1=12.0002 (7 characters)
after being received by the instrument,
and the last bit is unit code. When
resistance is measured, the nominal
value is 12 M
Ω
. That is, N1 starts with
number (the first one from 0~9), the
rest being numbers (0~9) and point,
and ends with unit code. The number of
total characters should be less than 7
after automatic adjustment.
2.
N2=12 will be automatically
adjusted to N2=12.000 (6 characters)
after being received by the instrument,
that is, N2 (Q/D limit value) starts with
number (the first one from 0~9), and
the rest are number 0~9 and point. The
number of total characters should be
less than 6 after automatic adjustment.
3.
N3=1% will be automatically
adjusted to N3=+1.000% (7
characters) after being received by the
instrument, that is, N3~N8 (limit
value) starts with symbols or numbers.
If starting with symbol, the second
character should be number, the rest
being number 0~9 and point, and ends
with %. The total characters should be
less than 7 after automatic adjustment.
2. N2: X1: ASCII code (0
∼
9);
X
2
∼
X
6
: ASCII code (0~9 and
point);
3. N3
∼
N8:
X
1
: ASCII code (+, -);
X
2
: ASCII code (0
∼
9);
X
3
∼
X
6
: ASCII code (0~9 and
point);
X
7
: ASCII code (%)