![Tonghui Electronics TH9201 Series Operation Manual Download Page 57](http://html1.mh-extra.com/html/tonghui-electronics/th9201-series/th9201-series_operation-manual_1126463057.webp)
◇
57
Command message:
SOUR:SAFE:STEP
<sn>:DC:CHAN
<channel>:
<HIGH/LOW/OPEN>
Query message:
SOUR:SAFE:STEP <sn>:DC: CHAN <channel>?
--Program data
<sn>:
Data format: Integer
Set value: 1~49
Resolution: 1
Unit:
--Program data
< channel >:
Data format: Integer
Set value: 1~8
Resolution: 1
Unit:
--Program data
<HIGH/LOW/OPEN>:
Data format: Character
Set value: HIGH/LOW/OPEN
Resolution:
Unit:
--
Example:
To set the scanner channel 1 for step 1 to HIGH,
:
SOUR:SAFE:STEP 1:DC:CHAN 1:HIGH.
--
Example:
SOUR:SAFE:STEP 1:DC:CHAN 1?
If the scanner channel 1 for step 1 is HIGH,
HIGH is returned
4.
Commands for IR Settings
:SOURce:SAFEty:STEP:IR:LEVel
Sets the test voltage for IR test.
Also inquires about the test voltage.
--Syntax
:
Command message: :
SOUR:SAFE:STEP <sn>:IR:LEV <volt>
Query message:
:SOUR:SAFE:STEP <sn>:IR:LEV?
--Program data
<sn>:
Data format: Integer