28
OFF
No contact test in low terminal.
KEY
Press ‘START’ to do contact test.
GR CONT
:
0.2S~99.9S
Contact time setup.
BEEP VOL
:
OFF
、
LOW
、
HIGHT Alarm volume.
CONTRAST
:
1
~
10
LCD contrast.
PASSWORD
:
ON
、
OFF
Key lock setup.
GFI
:
ON
、
OFF
Discharge wall function.
SYSTEM2
AFTR FAIL:
STOP
LOCK
RAMP JUDG:
OFF
OFST
DC50 AGC:
OFF
RMT
PART NO:
0000
ERR
SYS1
SYS2
INT
F1 F2 F3 F4 F5
4.2.5 SYSTEM2
Instruction
:
Label
Instruction
Definition
STOP
Use ‘STOP’ to quit in fail result.
CONTINUE
Continue the next step in fail result.
AFTR FAIL
:
RESTART
Retest in fail result.
RAMP JUDG
:
ON
、
OFF
Low limit judge in rising withstanding
voltage.
DC50 AGC
:
ON
、
OFF
High voltage hardware feedback in
DC50V-500V.
PART NO
:
8 digit
No. of test programme.
INTERFAC
GRIB ADD
:
03
DATA
:
8
LOCK
BAUD
:
19200
PARITY
:
NONE
OFST
STOP
:
2
RMT
ERR
SYS1
SYS2
INT
F1 F2 F3 F4 F5
Instruction
:
GRIB ADD
:
03 Communication
interface
address.
BAUD
:
19200
Communication interface baud rate..
STOP
:
2
Communication interface stop digit.
DATA
:
8
Communication interface data digit.
PARITY
:
NONE Communication
interface even odd check.