![Tonghui Electronics TH9201 Series Operation Manual Download Page 39](http://html1.mh-extra.com/html/tonghui-electronics/th9201-series/th9201-series_operation-manual_1126463039.webp)
◇
39
Sets the AUTO RANGE’s status.
Also inquires about the current AUTO RANGE’s status.
--Syntax
:
Command message:
:SYST:WRAN < ON/OFF/1/0 >
Query message:
:SYST:WRAN?
--Program data
<ON/OFF>:
Data format: Character
Set value: 0 (OFF) , 1 (ON)
--
Example:
To set the AUTO RANGE’s status to ON,
:SYST:WARN ON.
--
Example:
:SYST:WRAN?
If the current AUTO RANGE’s status is ON,
ON is returned.
:SYSTem:GCONtinuity
Sets the GR CONT’s status.
Also inquires about the current GR CONT’s status.
--Syntax
:
Command message:
:SYST:GCON < ON/OFF/1/0 >
Query message:
:SYST:GCON?
--Program data
<ON/OFF>:
Data format: Character
Set value: 0 (OFF) , 1 (ON)
--
Example:
To set the GR CONT’s status to ON,
:SYST:GCON ON.
--
Example:
:SYST: GCON?
If the current GR CONT’s status is ON,
ON is returned.
:SYSTem:GFI