17
3.2
Instruction of rear panel
WARNING:
FOR CONTINUED PROTECTION AGAINST FIRE
HAZARD,REPLACE ONLY WITH THE SAME TYPE AND RATING OF FUSE
AS SPECIFIED FOR THE LINE VOLTAGE BEING UTILIZED.
CAUTION:
NO OPERATOR SERVICEABLE PARTS INSIDE
REFER SERVICING TO QUALITIFIED PERSONNEL.
90V-110V~
108V-132V~
198V-242V~
216V-250V~
5AT
250V
2.5AT
250V
HIGH VOLTAGE
MAX:5kV AC 30mA/6kV DC 10mA
HV OUTPUT
RTN/
LOW
100V/120V/220V/240V~
50/60Hz 500W MAX
GPIB
CONT.
CHECK
OPTION
1
Ω
CAL
VOLTAGE SELECTOR
POLLUTION DEGREE 2
INSTALLATION CATEGORY Ⅱ
贴标签
!
!
OUTPUT
FAIL
START
RESET
COM
INTER
LOCK
UNDER
TEST
PASS
1
5
9
UNDER
RESET
TEST
START
PASS
FAIL
6
SCANNER
RS-232C
1
2
3
4
5
6
7
8
9
10
11
12
13
14
Figure 3-2
i. Fan
Power amplifier circuit radiator.
ii.
Line voltage range
Switch of input voltage range.
iii. PLC controller interface
Be used to connect programming controller, where
:
INTERLOCK
:
input the connecting locked signal
,
if off, starting output is not allowed.
TEST
:
Output synchronized-signal control when high voltage output is started.
START
:
Input the starting signal for outputting high voltage
,
corresponding to START
signal on the front panel.
RESET
:
PASS
:
FAIL
:
iv. OUTPUT
TEST outputs control
,
output 24V level or relay switch signal
,
it is suggested to use switch
signal to assure signal separation.
v. HANDEL
interface
Use 9 core model D jack to output.
vi.
Ground break-over test interface
When open ground break-over test function
,
the terminal has to be connected to the place
where DUT is connected with low terminal.
vii. SCAN
interface
Be used to connect TH9121 multi-channel tester.
viii. IEEE488
(
GPIB
)
parallel communication interface
Provide the general communication interface for instrument and external device.