4.3 Test item interface and parameter setup
4.3.1
AC withstanding voltage test parameter setup
:
STEP: 01/01
AC
SCAN:
1X2X3X4X5X6X7X8X
VOLT: 1.000 kV RISE: 0.5
S
LOC
UPPER: 1.000 mA FAIL: 0.5
S
OFST
REAL: 1.000
mA
LOW: 0.001
mA
RND
TIME: 1.0 S
FREQ:
50
Hz
ERR
ARC: 5.00
mA
T
S
F1 F2 F3 F4 F5
Figure 4.3.1
AC setup interface
Instruction
:
VOLT:
0.005~5.000kV
Voltage value of AC high voltage test
UPPER:
0.01~30.00mA
Current high limit value of AC
withstanding voltage
REAL:
OFF~0.001~30.00mA
Real current high limit value of AC
withstanding voltage
TIME:
OFF~0.1~999.9S
Test time of AC withstanding voltage
ARC:
OFF~0.1~15.0 mA
Current Max. value of AC arc
RISE:
OFF~0.1~999.9S
Voltage rising time of AC high voltage
test
FAIL:
OFF~0.1~999.9S
Voltage down time of AC high voltage
test
LOW:
OFF~0.001~30.00mA
Low limit current value of AC
FREQ:
50/60
AC working frequency
4.3.2
DC withstanding voltage test parameter setup
(
DC
):
STEP
:
01/01 DC SCAN: 1X2X3X4X5X6X7X8X
VOLT
:
1.000 kV RISE: 0.5
S
LOC
LOW
:
1.000 mA FAIL: 0.5
S
OFST
UPPER
:
1.000 mA WAT:
0.2
S
RND
TIME
:
1.0 S CHEK: OFF
ERR
ARC
:
5.00 mA
T
S
F1 F2 F3 F4 F5
Figure 4.3.2
DC setup interface
Instruction
:
VOLT:
0.050~6.000kV
Voltage value of DC high voltage test
UPPER:
0.001~10.00mA
Current high limit value of DC high
voltage
LOW
:
OFF~0.001~10.00mA
Current low limit value of DC
withstanding voltage
TIME:
OFF~0.1~999.9S
Test time of DC withstanding voltage
ARC:
OFF~0.1~10.0 mA
Current Max. value of DC arc