background image

 

 

TH9201 Series 

AC/DC Withstanding Voltage/ 

Insulation resistance Tester 

 
 
 

 

Operation  Manual 

 
 

 

First Edition 

Tonghui Electronic Co.,Ltd. 

www.tonghui.com.cn

 

 

 

Summary of Contents for TH9201 Series

Page 1: ... 1 TH9201 Series AC DC Withstanding Voltage Insulation resistance Tester Operation Manual First Edition Tonghui Electronic Co Ltd www tonghui com cn ...

Page 2: ...ensitive measuring 5 instrument or receiver 5 1 3 PRECAUTIONS FOR MOVING 6 1 3 1 Before moving the tester turn off the power switch 6 1 3 2 When moving the tester Disconnect all wires from it 6 1 4 CHECKING AND REPLACING FUSE 6 1 5 CONNECTING THE AC POWER CORD 7 1 6 GROUNDING 7 1 7 CHECKING OPERATIONS 8 1 8 OTHER SPECIFICATIONS 9 CHAPTER 2 PRECAUTIONS ON HANDLING 10 2 1 PROHIBITED OPERATIONS 10 2 ...

Page 3: ...D FUNCTIONS 15 3 1 FRONT PANEL 15 3 2 INSTRUCTION OF REAR PANEL 17 3 3 DESCRIPTION 19 CHAPTER 4 BASIC OPERATION 24 4 1 TURNING ON THE POWER 24 4 1 1 Turning on the power 24 4 2 INSTRUCTION OF INTERFACE FUNCTION 25 4 3 TEST ITEM INTERFACE AND PARAMETER SETUP 30 4 4 TEST FUNCTION THEORY AND INSTRUCTION 32 CHAPTER 5 TH9201 RS232 Commands 37 ...

Page 4: ...ltage test stick 1 TH26034 RS232C cable 1 AC Power cord 1 Operation Manual 1 Test Report 1 Servicing card 1 Accessories ordered by custom 1 1 2 Precautions for Installation Be sure to observe the following precautions when installing the tester 1 2 1 Do not use the tester in a flammable atmosphere To prevent explosion or fire do not use the tester near alcohol thinner or other combustible material...

Page 5: ...vide sufficient space for the air inlet on the lateral side and the air outlet on the rear side to allow air to flow 1 2 7 Do not place the tester on a tilted surface or in a location subject to vibrations If placed on a non level surface or in a location subject to vibration the tester may fall resulting in damage and injury 1 2 8 Do not use the tester in locations affected by strong magnetic or ...

Page 6: ...he fuse and the source voltage agree with the LINE VOLTAGE RANGE switch on the rear panel Nominal voltage range allowable voltage range 100 V to 120 V AC 85 V to 132 V AC 200 V to 240 V AC 170 V to 250 V AC Allowable frequency range 47 Hz to 63 Hz WARNING To prevent malfunctions be sure to operate within the line voltage range 1 4 Checking and replacing fuse WARNING To prevent electric shock befor...

Page 7: ...m that the POWER switch on the tester is off 3 Connect the AC power cord to the AC LINE connector on the rear panel Use the provided power code or power code that is selected by qualified personnel 4 Plug in the AC power cord 1 6 Grounding WARNING Be sure to connect the tester to an electrical ground safety ground If the output to a conveyer or peripheral device that is connected to an earth groun...

Page 8: ... off every time the door is opened Before turning on the power confirm that the allowable voltage range indicated on the power supply is the same as that indicated on the rear panel of the tester When the power is turned on the tester lights all LEDs on the front panel and self diagnosis is started Before starting up the tester confirm that all LEDs are on to ensure safety It is particularly dange...

Page 9: ...e POWER switch 6 Turn on the POWER switch again RN 7 Following the opening screen display the ACW screen and confirm that the tester is kept in the READY status The above steps complete the checking procedure Opening screen 同 惠 电 子 T H 9 2 0 1 VERSION 1 0 1 8 Other specifications 1 Power 500VA TH9201 S 350VA TH9201B C 2 Dimensions 330mm 140mm 385mm 3 Weight 20kg TH9201 S 16kg TH9201B C ...

Page 10: ...tion so that the high test voltage line is not shorted to a nearby AC line or nearby devices such as conveyors which are connected to an earth ground If it is shorted the tester chassis can be charged up to the hazardous high voltage Be sure to connect the protective grounding terminal of the tester to an earth line If this has been securely done even when the HIGH VOLTAGE terminal is shorted to t...

Page 11: ...d output terminals and their vicinities can be charged up to dangerous high voltages Never touch them when in test WARNING The vinyl sheaths of the alligator clips of the test leadwires which are supplied accompanying the tester have no sufficient insulation for the high test voltages Never touch them when in test 2 3 4 Matters to be Sure of After Turning OFF Power If you have to touch the DUT tes...

Page 12: ... the following lengths of time to reduce the internal capacitor voltage to 30 V If the DUT is disconnected during a test or before the completion of discharging assuming that the DUT has a capacity of 0 01uF and a parallel resistance of 100 M approximately 5 3 seconds at 5 kV and approximately 3 5 seconds at 1 kV are required for the DUT to discharge to 30 V When the approximate time constant of t...

Page 13: ...nges the output block may be heated to excess activating the internal protection circuit Should this happen wait until the temperature returns to the normal level Output requirements for withstanding voltage testing Ambient temperature Upper current Pause Time Output time 20mA TH9201 S 12mA TH9201B C At least as long as the output time Maximum of 1 minute AC 10mA TH9201 S 6mA TH9201B C Not necessa...

Page 14: ...ng of the high voltage test lead wire is free from cracks fissures and breakage 3 The high voltage test lead wire is not broken 4 The tester generates FAIL signal when the ends of the low voltage test lead wire and high voltage test lead wire are short circuited ...

Page 15: ...d STOP Start up the instrument once test starts TEST will be displayed at the left corner of the screen TEST indicator lights Stop key used to cancel the test Or PASS FAIL status 2 Band and model 3 LCD screen 240 64 dot matrix LCD screen display setting and test interface etc 4 FUNCTION Select mode system interface TEST Press the key and the corresponding key lights the instrument is ready to test...

Page 16: ...splay contrast 7 Code switch When in READY status set the test condition on LCD 8 USB interface Externally connect to USB storage 9 Output voltage HIGH and LOW terminal High and low voltage terminal in test voltage output Warning In the process of testing don t touch the high terminal Caution if external voltage is used in test terminal it will cause the damage of internal circuit 10 Shortcut key ...

Page 17: ...Switch of input voltage range iii PLC controller interface Be used to connect programming controller where INTERLOCK input the connecting locked signal if off starting output is not allowed TEST Output synchronized signal control when high voltage output is started START Input the starting signal for outputting high voltage corresponding to START signal on the front panel RESET PASS FAIL iv OUTPUT...

Page 18: ...ack Be used to input AC power please use the attached power line with fuse Change the fuse according to input power xii Ground break over calibration Ground break over resistance is 1 ohm use potentiometer to calibrate if there is deviation xiii Mark This mark describes the instrument s SN xiv High voltage output interface optional Stand by high voltage output interface ...

Page 19: ... use 600Hz AC to form DC voltage to be the power which can assure the DC power ripple is far less than the formal withstanding voltage tester thus to assure the stability of 5000V 10mA test Once connecting the load TH9201 not only can perform the independent AC withstanding test DC withstanding voltage test insulation resistance test but also multi item test via the setting of test programme When ...

Page 20: ...essary to adjust output voltage after setting test voltage Figure 3 3 AC voltage load adjusting rate DC withstanding voltage test 6kV Max output power is 50W TH9201 series can provide DC withstanding voltage test of wide voltage range Max output power is 50W the max holding time is 1 min Instrument allocates a reliable low ripple DC AC switch circuit voltage load adjusting rate 1 Figure 3 4 DC vol...

Page 21: ...thstanding voltage test and insulation resistance test the test voltage can be raised to set value slowly it can t provide set voltage to DUT instantly after test starts The resolution is 0 1s from 0 1s to 99 9s and the resolution is from 100s to 200s TH9201series meet each UL test standard and IEC withstanding voltage standard primary voltage is less than the half of test voltage and it can appoi...

Page 22: ... rising time control function Easy operation TH9201series are quite easy to be operated In the setup interface the test condition is listed Set test condition use direction key to select a topic from LCD display then rotate code switch Shortcut key can select the set topic After setting the data it will back to test interface to perform test 50 test programmes and each one includes 100 test items ...

Page 23: ...ARNING The instrument applies 5kV AC DC high voltage so do not touch the DUT and test line or it will cause the danger of electric shock The safe measures should be taken around DUT to assure the operator s safety Otherwise please be care about the outputted high voltage due to the incorrect connection and operation ...

Page 24: ...re starting up the tester that all LEDs are lit It is particularly dangerous to start a test when the DANGER lamp is broken Even when the DANGER lamp is lit no output or voltage is being generated CAUTION When the POWER switch has been turned off wait several seconds before turning it on again Turning the POWER switch on off repeatedly at insufficient intervals may damage the tester CAUTION Even a...

Page 25: ...uction of operation Switch the current test item from default AC withstanding voltage AC to open detection OS The operation is as below Press X underline to AC see the figure below The following indication is displayed in F1 F5 Now press F4 in the figure below the position of AC will be changed to OS the following item parameter will also be changed to default one of open detection Clockwise rotat...

Page 26: ...e Cursor in this position can switch to other working items by F1 F3 for instance DC IR OS SCAN 1X2X3X4X5X6X7X8X Connect multi channel output and test terminal SCAN channel as for TH9201S with multi channel other models are corresponded with output terminal of TH9121multi channelstester 8 output channels is connected with test output terminal there are open high voltage output and ground low termi...

Page 27: ...t condition of DUT with the unit of S Note operator can t leave if the tester is working and in the process of testing not close to the test line or DUT 4 2 3 SYSTEM SYSTEM1 PASS HOLD 0 5 S BEEP VOL LOW LOCK STEP HOLD 0 5 S CONTRAST 4 OFST AUTO RANG OFF PASSWORD OFF RMT GR CONT 0 2 S GFI OFF ERR SYS1 SYS2 INT F1 F2 F3 F4 F5 4 2 4 SYSTEM1 Instruction Label Instruction Definition PASS HOLD 0 2S 99 9...

Page 28: ...STOP to quit in fail result CONTINUE Continue the next step in fail result AFTR FAIL RESTART Retest in fail result RAMP JUDG ON OFF Low limit judge in rising withstanding voltage DC50 AGC ON OFF High voltage hardware feedback in DC50V 500V PART NO 8 digit No of test programme INTERFAC GRIB ADD 03 DATA 8 LOCK BAUD 19200 PARITY NONE OFST STOP 2 RMT ERR SYS1 SYS2 INT F1 F2 F3 F4 F5 Instruction GRIB A...

Page 29: ...T REAL 1 000 mA NAME UNNAME RMT TIME 1 0 S STEP Black ERR ARC 5 00 mA T S ENTER F1 F2 F3 F4 F5 Instruction Label Instruction Definition SN 00 49 SN of memory NAME 8 digit Name of programme Black The current memory is not available STEP 00 99 The number of item saved in memory Operator selects the memory and fill in the file name press F4 ENTER to enter dialog box to confirm Save data Load interfac...

Page 30: ... Current Max value of AC arc RISE OFF 0 1 999 9S Voltage rising time of AC high voltage test FAIL OFF 0 1 999 9S Voltage down time of AC high voltage test LOW OFF 0 001 30 00mA Low limit current value of AC FREQ 50 60 AC working frequency 4 3 2 DC withstanding voltage test parameter setup DC STEP 01 01 DC SCAN 1X2X3X4X5X6X7X8X VOLT 1 000 kV RISE 0 5 S LOC LOW 1 000 mA FAIL 0 5 S OFST UPPER 1 000 m...

Page 31: ... 10 00mA High limit value of IR LOW OFF 0 001 10 00mA Low limit value of IR TIME OFF 0 1 999 9S Test time of IR RANG AUTO 300nA 3uA 30uA 300uA 3mA 10mA Range control of IR RISE OFF 0 1 999 9S Rising time of insulation voltage FAIL OFF 0 1 999 9S Down time of insulation voltage SAGC ON OFF Software auto voltage control 4 3 4 Open detection OS parameter setup STEP 1 1 OS SCAN 1X2X3X4X5X6X7X8X OPEN C...

Page 32: ... value but the value of the sampled current being transferred via impedance The value should be close to that of capacitor installed in the test terminal The sampling current is not only generated by capacitor 4 4 Test function theory and instruction Start test High voltage test Voltage down Ground test Test voltage rise Charge current test Electric wall Pass judge End Arc detection Current limit ...

Page 33: ...ion is normal within the set time if circle resistance 1ohm it is normal the ground check is passed and test continues If test circuit is not available if circle resistance 1ohm it is abnormal the instrument will quit testing and displays connection error GR FAIL Note ground breakover parameter is the GR CONT in SYSTEM1 interface 1 Electric wall function Electric wall is used to check the ground c...

Page 34: ...imit judge REAL DUT is considered as capacitive and the capacitor takes current in AC test When the capacitive change is big then the capacitive current has nothing to do with the test result now you can select real current judge which only judges the current of same voltage phase It displays REAL FAIL in over limit ARC It is a practical function in component test which tests the instant fire in p...

Page 35: ...ir breakdown in high voltage which causes the partial high frequency self motivation There is low frequency current in current sampling circuit and the fast signal can t be handled meanwhile Arc detection circuit only samples the changeable amplitude to find the potential defect in current return circuit ...

Page 36: ... after test ends 8 channels sweep Available Voltage display Test range 0 00kV 6 00kV AC DC Digit Accuracy 1 0 reading 2V Current display AC 0 1mA 30mA 0 1mA 20mA Test range DC 0 01mA 10mA 0 01mA 5mA Accuracy 1 reading 10digits after correction AC 1mA 15mA 1mA 15mA Arc detecti on DC 1mA 10mA 1mA 5mA Insulation resistance display Resistance test range 0 01MΩ 9 99GΩ current range is from10nA 10mA Res...

Page 37: ... 1 00 is returned SYSTem TIME PASS Sets the pass hold time in a PASS judgment Also inquires about the current PASS hold time Syntax Command message SYST TIME PASS time Query message SYST TIME PASS Program data Data format float Set value 0 3 99 9 Resolution 0 1 Unit s Example To set the PASS hold time to 1 0s SYST TIME PASS 1 0 Example SYST TIME PASS If the current PASS hold time is 1 0 1 0 is ret...

Page 38: ... TIME STEP 1 0 Example SYST TIME STEP If the current step hold time is 1 0 1 0 is returned SYSTem RJUDgment Sets the RAMP JUDG s status Also inquires about the current RAMP JUDG s status Syntax Command message SYST RJUD ON OFF 1 0 Query message SYST RJUD Program data ON OFF Data format Character Set value 0 OFF 1 ON Example To set the RAMP JUDG to ON SYST RJUD ON Example SYST RJUD If the current R...

Page 39: ... to ON SYST WARN ON Example SYST WRAN If the current AUTO RANGE s status is ON ON is returned SYSTem GCONtinuity Sets the GR CONT s status Also inquires about the current GR CONT s status Syntax Command message SYST GCON ON OFF 1 0 Query message SYST GCON Program data ON OFF Data format Character Set value 0 OFF 1 ON Example To set the GR CONT s status to ON SYST GCON ON Example SYST GCON If the c...

Page 40: ...T GFI If the current GFI s status is ON ON is returned SYSTem FAIL Sets the AFTR FAIL s status Also inquires about the current AFTR FAIL s status Syntax Command message SYST FAIL STOP CONTinue RESTart Query message SYST FAIL Program data STOP CONTinue RESTart Data format Character Set value STOP CONTinue RESTart Example To set the AFTR FAIL s status to STOP SYST FAIL STOP Example SYST FAIL If the ...

Page 41: ...er volume to 1 SYST BEEP 1 Example SYST BEEP If the current buzzer volume is 1 1 is returned SYSTem DAGC Sets the DC50 AGC s status Also inquires about the current DC50 AGC s status Syntax Command message SYST DAGC ON OFF 1 0 Query message SYST DAGC Program data ON OFF Data format Character Set value 0 OFF 1 ON Example To set the DC50 AGC s status to ON SYST DAGC ON Example SYST DAGC If the curren...

Page 42: ...rmat Integer Set value 1 10 Resolution 1 Unit Example To set the LCD contrast to 4 SYST CR 4 Example SYST CR If the current LCD contrast is 4 4 is returned SYSTem PART Sets the part number Also inquires about the current part number Syntax Command message SYST PART set value Query message SYST PART Program data Data format Integer Set value 8 位 Resolution Unit Example To set the part number to 200...

Page 43: ...data ON OFF Data format Character Set value 0 OFF 1 ON Example To set the KEY LOCK s status to ON SYST LOCK ON Example SYST LOCK If the current KEY LOCK s status is ON ON is returned 2 Commands for AC settings SOURce SAFEty STEP AC LEVel Sets the test voltage for ACW test Also inquires about the test voltage Syntax Command message SOUR SAFE STEP sn AC LEV voltage Query message SOUR SAFE STEP sn AC...

Page 44: ...P AC LIMit LOW Sets the LOWER current for ACW test Also inquires about the LOWER current Syntax Command message SOUR SAFE STEP sn AC LIM LOW current Query message SOUR SAFE STEP sn AC LIM LOW Program data sn Data format Integer Set value 1 49 Resolution 1 Unit Program data current Data format float Set value 0 30 000E 3 0 is OFF Resolution 1 000E 6 Unit A Example To set the LOWER current for step ...

Page 45: ...e 1 49 Resolution 1 Unit Program data current Data format float Set value 1 00E 6 30 000E 3 Resolution 1 000E 6 Unit A Example To set the UPPER current for step 1 to 0 001 SOUR SAFE STEP 1 AC LIM LOW 0 001 Example SOUR SAFE STEP 1 AC LIM LOW If the current UPPER current for step 1 is 0 001 0 001 is returned SOURce SAFEty STEP AC LIMit ARC Sets the ARC current for ACW test Also inquires about the A...

Page 46: ...rrent ARC current for step 1 is 0 001 0 001 is returned SOURce SAFEty STEP AC LIMit REAL Sets the REAL current for ACW test Also inquires about the REAL current Syntax Command message SOUR SAFE STEP sn AC LIM REAL current Query message SOUR SAFE STEP sn AC LIM REAL Program data sn Data format Integer Set value 1 49 Resolution 1 Unit Program data current Data format float Set value 0 30 000E 3 0 is...

Page 47: ...FE STEP sn AC TIME RAMP Program data sn Data format Integer Set value 1 49 Resolution 1 Unit Program data time Data format float Set value 0 999 9 0 is OFF Resolution 0 1 Unit s Example To set the RISE time for step 1 to 1 SOUR SAFE STEP 1 AC TIME RAMP 1 Example SOUR SAFE STEP 1 AC TIME RAMP If the current RISE time for step 1 is 1 1 is returned SOURce SAFEty STEP AC TIME FALL Sets the FALL time f...

Page 48: ...ALL 1 Example SOUR SAFE STEP 1 AC TIME FALL If the current FALL time for step 1 is 1 1 is returned SOURce SAFEty STEP AC TIME TEST Sets the test time for ACW test Also inquires about the test time Syntax Command message SOUR SAFE STEP sn AC TIME TEST time Query message SOUR SAFE STEP sn AC TIME TEST Program data sn Data format Integer Set value 1 49 Resolution 1 Unit Program data time Data format ...

Page 49: ...UR SAFE STEP sn AC TIME FREQ Program data sn Data format Integer Set value 1 49 Resolution 1 Unit Program data freq Data format Character Set value 50 60 Resolution Unit Hz Example To set the test frequency for step 1 to 50 SOUR SAFE STEP AC TIME FREQ 50 Example SOUR SAFE STEP AC TIME FREQ If the current test frequency for step 1 is 50 50 is returned SOUR SAFE STEP AC CHAN Sets HIGH LOW OPEN for t...

Page 50: ...W OPEN Resolution Unit Example To set the scanner channel 1 for step 1 to HIGH SOUR SAFE STEP 1 AC CHAN 1 HIGH Example SOUR SAFE STEP 1 AC CHAN 1 If the scanner channel 1 for step 1 is HIGH HIGH is returned 3 Commands for DC settings SOURce SAFEty STEP DC LEVel Sets the test voltage for DCW test Also inquires about the test voltage Syntax Command message SOUR SAFE STEP sn DC LEV volt Query message...

Page 51: ...d SOURce SAFEty STEP DC LIMit LOW Sets the LOWER current for DCW test Also inquires about the LOWER current Syntax Command message SOUR SAFE STEP sn DC LIM LOW current Query message SOUR SAFE STEP sn DC LIM LOW Program data sn Data format Integer Set value 1 49 Resolution 1 Unit Program data current Data format float Set value 0 10 000E 3 0 is OFF Resolution 1 000E 6 Unit A Example To set the LOWE...

Page 52: ...data sn Data format Integer Set value 1 49 Resolution 1 Unit Program data current Data format float Set value 1 00E 6 10 000E 3 Resolution 1 000E 6 Unit A Example To set the UPPER current for step 1 to 0 001 SOUR SAFE STEP 1 DC LIM LOW 0 001 Example SOUR SAFE STEP 1 DC LIM LOW If the current UPPER current for step 1 is 0 001 0 001 is returned SOURce SAFEty STEP DC LIMit ARC Sets the ARC current fo...

Page 53: ...C LIM ARC 0 001 Example SOUR SAFE STEP 1 DC LIM ARC If the current ARC current for step 1 is 0 001 0 001 is returned SOUR SAFE STEP DC TIME RAMP Sets the RISE time for DCW test Also inquires about the RISE time Syntax Command message SOUR SAFE STEP sn DC TIME RAMP time Query message SOUR SAFE STEP sn DC TIME RAMP Program data sn Data format Integer Set value 1 49 Resolution 1 Unit Program data tim...

Page 54: ... sn DC TIME FALL time Query message SOUR SAFE STEP sn DC TIME FALL Program data sn Data format Integer Set value 1 49 Resolution 1 Unit Program data time Data format float Set value 0 999 9 0 is OFF Resolution 0 1 Unit s Example To set the FALL time for step 1 to 1 SOUR SAFE STEP 1 DC TIME FALL 1 Example SOUR SAFE STEP 1 DC TIME FALL If the current FALL time for step 1 is 1 1 is returned SOUR SAFE...

Page 55: ... 1 to 1 SOUR SAFE STEP 1 DC TIME TEST 1 Example SOUR SAFE STEP 1 DC TIME TEST If the current TEST time for step 1 is 1 1 is returned SOUR SAFE STEP DC TIME DWEL Sets the WAIT time for DCW test Also inquires about the WAIT time Syntax Command message SOUR SAFE STEP sn DC TIME DWEL time Query message SOUR SAFE STEP sn DC TIME DWEL Program data sn Data format Integer Set value 1 49 Resolution 1 Unit ...

Page 56: ... CLOW ON OFF or 1 0 Query message SOUR SAFE STEP sn DC CLOW Program data sn Data format Integer Set value 1 49 Resolution 1 Unit Program data set value Data format Character Set value OFF 0 ON 1 Resolution Unit Example To set the CHECK LOW s status for step 1 to ON SOUR SAFE STEP DC CLOW ON Example SOUR SAFE STEP DC CLOW If the current CHECK LOW s status for step 1 is ON ON is returned SOUR SAFE S...

Page 57: ...ormat Character Set value HIGH LOW OPEN Resolution Unit Example To set the scanner channel 1 for step 1 to HIGH SOUR SAFE STEP 1 DC CHAN 1 HIGH Example SOUR SAFE STEP 1 DC CHAN 1 If the scanner channel 1 for step 1 is HIGH HIGH is returned 4 Commands for IR Settings SOURce SAFEty STEP IR LEVel Sets the test voltage for IR test Also inquires about the test voltage Syntax Command message SOUR SAFE S...

Page 58: ... returned SOURce SAFEty STEP IR LIMit LOW Sets the LOWER resistance for IR test Also inquires about the LOWER resistance Syntax Command message SOUR SAFE STEP sn IR LIM LOW resistance Query message SOUR SAFE STEP sn IR LIM LOW Program data sn Data format Integer Set value 1 49 Resolution 1 Unit Program data resistance Data format float Set value 1 0E5 5 0E10 Resolution 1 0E5 Unit Ω Example To set ...

Page 59: ...H Program data sn Data format Integer Set value 1 49 Resolution 1 Unit Program data resistance Data format float Set value 0 5E10 0 is OFF Resolution 1 0E5 Unit Ω Example To set the UPPER resistance for step 1 to 1000000 SOUR SAFE STEP 1 IR LIM HIGH 1000000 Example SOUR SAFE STEP 1 IR LIM HIGH If the current UPPER resistance for step 1 is 1000000 1000000 is returned SOUR SAFE STEP IR TIME RAMP Set...

Page 60: ... RAMP 1 Example SOUR SAFE STEP 1 IR TIME RAMP If the current RISE time for step 1 is 1 1 is returned SOUR SAFE STEP IR TIME FALL Sets the FALL time for IR test Also inquires about the FALL time Syntax Command message SOUR SAFE STEP sn IR TIME FALL time Query message SOUR SAFE STEP sn IR TIME FALL Program data sn Data format Integer Set value 1 49 Resolution 1 Unit Program data time Data format flo...

Page 61: ...STEP sn IR TIME TEST time Query message SOUR SAFE STEP sn IR TIME TEST Program data sn Data format Integer Set value 1 49 Resolution 1 Unit Program data time Data format float Set value 0 999 9 0 is OFF Resolution 0 1 Unit s Example To set the TEST time for step 1 to 1 SOUR SAFE STEP 1 IR TIME TEST 1 Example SOUR SAFE STEP 1 IR TIME TEST If the current TEST time for step 1 is 1 1 is returned SOUR ...

Page 62: ...ON SOUR SAFE STEP IR AGC ON Example SOUR SAFE STEP IR AGC If the current AGC s status for step 1 is ON ON is returned SOUR SAFE STEP IR CHAN Sets HIGH LOW OPEN for the scanner channel for IR test Also inquires about the set value for the scanner channel Syntax Command message SOUR SAFE STEP sn IR CHAN channel HIGH LOW OPEN Query message SOUR SAFE STEP sn IR CHAN channel Program data sn Data format...

Page 63: ...channel 1 for step 1 is HIGH HIGH is returned 5 Commands for OS Settings SOUR SAFE STEP OSC OPEN Sets the OPEN rate for OS test Also inquires the OPEN rate Syntax Command message SOUR SAFE STEP sn OSC OPEN set value Query message SOUR SAFE STEP sn OSC OPEN Program data sn Data format Integer Set value 1 49 Resolution 1 Unit Program data set value Data format float Set value 0 1 1 0 Resolution 0 1 ...

Page 64: ... data sn Data format Integer Set value 1 49 Resolution 1 Unit Program data set value Data format Integer Set value 0 5 0 is OFF Resolution 1 Unit Example To set the SHORT rate for step 1 to 200 SOUR SAFE STEP 1 OS SHOR 2 Example SOUR SAFE STEP 1 OS SHOR If the SHORT rate for step 1 is 200 2 is returned SOUR SAFE STEP OS CHAN Sets HIGH LOW OPEN for the scanner channel for OS test Also inquires abou...

Page 65: ...cter Set value HIGH LOW OPEN Resolution Unit Example To set the scanner channel 1 for step 1 to HIGH SOUR SAFE STEP 1 OS CHAN 1 HIGH Example SOUR SAFE STEP 1 OS CHAN 1 If the scanner channel 1 for step 1 is HIGH HIGH is returned 6 Other Commands SOUR SAFE START Starts testing Syntax Command message SOUR SAFE START Example SOUR SAFE START SOUR SAFE STOP Stops a test Also cancels FAIL PASS STOP stat...

Page 66: ...ecall memory 1 SOUR SAFE LOAD 1 SOUR SAFE STEPSN Inquires about the current STEP during a test Syntax SOUR SAFE STEPSN Example SOUR SAFE STEPSN If the current STEP during a test is 2 2 is returned IDN Inquires about the tester s name and ROM version Syntax Command message IDN Example IDN TH9201 Ver 1 0 is returned TEST DATAI Inquires about the monitor current during the ACW DCW test Syntax TEST DA...

Page 67: ...turned TEST FETCH Inquires the test results of per step Syntax TEST FETCH Response message data1 data 2 datan Unit mA MΩ Example TEST FETCH If the result of step1 is 1 00mA the step2 s is 3 00mA 1 00 3 00 is returned TEST FETCH2 Inquires the test voltage current for ACW or DCW test or resistance for IR test status of the current step Syntax TEST FETCH2 Response message Status value voltage value c...

Page 68: ... Data format Integer Resolution 1 Unit V Program data current or resistance value Data format float Unit mA MΩ Example TEST FETCH2 If the current monitor voltage is 1000V the monitor current is 1 00mA in an ACW test 1 1000 1 00 is returned ...

Reviews: