![Tomy MX-107 Operator'S Manual Download Page 61](http://html1.mh-extra.com/html/tomy/mx-107/mx-107_operators-manual_1126195061.webp)
— 60 —
11-2. Cooling Capacity
<
Rotor cooling capacity
>
The MX-307, 207 and 107 are designed to perform the cooling capacity shown in the table below
at ambient temperatures of 25℃ or lower.
Rotors
Speed at the
operating
temperature of
4℃
Speed at the
operating
temperature of
10℃
TMP-21
15,000rpm
←
TMP-24
13,500rpm
←
TMA-100
15,000rpm
※
1
15,000rpm
TMA-200
15,000rpm
※
2
15,000rpm
※
4
TMA-300
15,000rpm
※
3
15,000rpm
※
5
TMA-22
15,000rpm
←
TMA-25BH
15,000rpm
←
TMA-26
16,000rpm
←
TMA-27
15,000rpm
←
TMA-29
15,000rpm
←
TMA-30
15,000rpm
←
TMA-32
15,000rpm
←
TMS-21
15,000rpm
←
※ 1 :
<MX-107>
・
When using the AR002-64 rack, the speed at the
operating temperature of 4℃ is 13,000rpm.
<MX-207>
・
When using the AR002-64 rack, the speed at the
operating temperature of 4℃ is 14,000rpm.
<MX-307>
・
When using the AR002-64 rack, the speed at the
operating temperature of 4℃ is 14,000rpm.
※ 2 :
<MX-207>
・
When using the AR002-64 rack, the speed at the
operating temperature of 4℃ is 13,500rpm.
<MX-307>
・
When using the AR002-64 rack, the speed at the
operating temperature of 4℃ is 14,000rpm.
※ 3 :
<MX-307>
・
When using the AR002-64 rack, the speed at the
operating temperature of 4℃ is 14,000rpm.
※ 4 :
<MX-207>
・
When using the AR002-64 rack, the speed at the
operating temperature of 10℃ is 14,000rpm.
<MX-307>
・
When using the AR002-64 rack, the speed at the
operating temperature of 10℃ is 14,500rpm.
※ 5 :
<MX-307>
・
When using the AR002-64 rack, the speed at the
operating temperature of 10℃ is 14,500rpm.
The chamber temperature may not be controlled
under the following conditions:
When the rotor stops spinning.
While accelerating or decelerating.
When running a short spin.
When spinning at low speed.
When the set temperature is higher than
ambient temperature.
・
・
・
・
・
<
Chamber temperature control
>