— 44 —
7-1. Setting the Rotational Radius(F0)
This function allows setting the value of the rotational radius. When the radius has been
changed, the revised RCF value will be calculated from the changed radius.
※
Factory default value is “81”.
※
For the rotational radius in each rotor, refer to “11-4. Rotor Specifications and List of Applicable
Tubes”.
Set value [mm]
Description
40
〜
87
The rotational radius
7-2. Setting the Adaptive Speed for the Soft Brake(F1)
This function allows setting the adaptive speed for the Soft Brake. When the power is turned
off, the set speed must be initialized again and returns to 500rpm.
Set speed [×100rpm]
Description
5
〜
150 (MX-107)
The soft brake is applied at a speed less than or equal to
the set speed.
5
〜
160 (MX-207, MX-307)
※
When turning on the unit again after turning off the power, the setting returns to “5”.
7-3. Selecting the Flash Mode Pattern(F2)
This function allows selecting the Flash mode pattern.
※
Factory default value is "1".
Set value
Description
0
Flash run continues as long as the flash key is held down.
1
The rotor accelerates in Flash run mode. Once it reaches to a set
speed, deceleration starts and allows the rotor to stop.
7-4. Setting the Speed in the Flash Mode(F3)
This function allows setting the speed in the Flash mode. In the actual flash mode, the lower
value of either the Flash speed or the speed set in the normal mode is applied.
※
Factory default value is "50".
Set speed [×100rpm]
Description
3
〜
150 (MX-107)
Flash speed
3
〜
160 (MX-207, MX-307)