— 28 —
The rotor rack adapter ARA-01 is designed for adjusting the level of a rack installed in the TMA-
300 Rotor so that it provides easy access to the rack.
Compatible racks for the ARA-01 adapter : AR015-24, AR002-64, and AR005-24
<
Rotor rack adapter ARA-01
>
After setting the rotor-rack adapter ARA-01 in the rotor TMA-300.
ARA-01 applicable rack
ARA-01
TMA-300
※
Do not use the ARA-01 adapter for other applications or in a way other than specified.
Installing the rotor-rack adapter ARA-01
When centrifuging microtubes with their caps being opened, install the rack properly into the
rotor in accordance with the following.
●
Do not stack the rack in piles.
●
Do not install the rack together with the ARA-01 adapter.
<
To install the rack with a tube cap being opened
>
※
If the rack accommodating the tubes with their caps being opened is installed in piles or with the
adapter, it may stick outside the rotor or may not be seated horizontally. Or the tube caps may touch
the chamber lid during spinning.
Installing a rack loaded with tubes with caps open
The rack alone should be installed.
TMA-100
TMA-200
TMA-300
TMA-200
TMA-300