— 13 —
3-1. Moving and Installing the Device
3. INSTALLATION
WARNING
Do not install the MX-307, 207 and 107 near chemical
storage facilities or sites likely to generate
flammable gases or fumes.
●
Electrical components can corrode and may cause a short
circuit or electrical shock.
Avoid installing the instrument in a place where
there is dust or moisture.
●
Failure to do so may cause a short circuit of electrical components or fire.
Avoid installing the instrument in or near a sink or tap where
there is a risk of it being splashed with water.
●
Failure to do so may cause a short circuit or electrical shock.
Install the instrument on a flat and level surface such as a concrete
floor.
●
Failure to do so may cause the instrument to fall down, thereby resulting in
injury to laboratory personnel or damage to the product and surrounding
property.
CAUTION
Do not move the MX-307, 207 and 107 with a rotor being mounted.
●
Otherwise, the motor shaft may bend, thereby causing the rotor to fall off
during spinning.