— 16 —
3-2. Connecting to the Power Source and Grounding
WARNING
Connect the MX-307, 207 and 107 to a power outlet of the voltage
specified as correct for the instrument. Do not connect the
instrument to a shared outlet.
●
It may cause an accidental fire, short circuit or malfunction.
Do not use extension cords or power strips.
●
It may cause an accidental fire, short circuit or
malfunction.
Ensure that the MX-307, 207 and 107 is properly
grounded.
●
Failure to do so may cause an explosion, electrical
shock or malfunction.
Never ground the MX-307, 207 and 107 through
a water main or gas pipe.
●
Such grounding may cause an explosion, electrical
shock or malfunction.
The MX-307, 207 and 107 should be properly grounded, thereby protecting the operator against
electrical shock from the power outlet. If the necessary grounding is not provided in your facility,
contact a qualified electrician for arrangement of a proper grounding.
Protection against electrical shock is provided by connecting the protective grounding wire to the
ground terminal.
Connect the power plug to the proper outlet. Be sure to provide protective grounding.
15A/125V
※
Unplug the power plug from the outlet when hearing thunder in order to prevent
lightning damage.
<Sample>
Rated voltage
Power Requirements
110/120V
1 - Phase AC110/120V 50/60Hz 15A
220/230/240V
1 - Phase AC220/230/240V 50/60Hz 8A