5-5-3. Swing-Out Rotor
CAUTION
Ensure that the same type of buckets are
correctly mounted into all four positions on
the swing-out rotor.
●
Otherwise, it may apply excessive force to the
rotor, thereby causing damage to the rotor.
Ensure that all buckets on the swing-out
rotor are allowed to move freely after loading
sample tubes.
●
Otherwise, unbalanced load may occur during
spinning, applying excessive force to the motor
shaft, thereby causing the rotor to fall off.