Operation Manual
44
8. TROUBLESCHOOTING
All possible failures are listed in table.
No
Failure
Couse
Action
1
TCP-2000 Fails to operate
when connected to the
mains ( Power LED is
OFF, no BEEP sound after
power ON )
Power cord broken or
power cord not connected
or no voltage in mains.
-Mother board or power
supply defective.
-Check power cord
-Check power cord connection
-Check main voltage in wall socket.
- Call service.
2
Device fails to operate
( Power LED is flashing
durin power-up process
and then stays ON )
-LCD backlight broken
-LCD screen broken or
control electronics broken
- Call service.
- Call service.
3
IR remote fails to operate
- Faulty battery
- IR remote broken
- IR receiver faulty
- Mother board broken
- Check and replace battery
- Replace IR remote
- Call service.
- Call service.
4
Message: : contact service
– Error 10.1
Turn off Chart Panel by pulling of
power plug. Wait two seconds and
plug in power cord. If problem
remains – call service.
5
Message: : contact service
– Error 10.2
Turn off Chart Panel by pulling of
power plug. Wait two seconds and
plug in power cord. If problem
remains – call service.
6
Message: : contact service
– Error 10.3
Turn off Chart Panel by pulling of
power plug. Wait two seconds and
plug in power cord. If problem
remains – call service.
Summary of Contents for TCP-2000