Operation Manual
13
4.1. Distance
This option allows user to set proper distance between patient and TCP-2000 Chart
Panel. It is crucial that this parameter is correctly set before first use of the Chart
Panel.
1.
Enter ‘
Distance
’, using Arrow Up or Arrow Down button.
2.
Press Right Arrow to enable distance edit mode.
3.
Use Up and Down Arrow to set required distance.
4.
To store value press Left Arrow.
4.2. AMD Distance
This option allows user to set proper distance between patient with
AMD
(Age-Related Macular Degeneration)
and TCP-2000 Chart Panel.
1.
Enter ‘
AMD Distance
’, using Arrow Up or Arrow Down button.
2.
Press Right Arrow to enable distance edit mode.
3.
Use Up and Down Arrow to set required distance.
4.
To store value press Left Arrow.
It is possible to set the distance from 0,5m – 1,8m.
4.3. AMD Optotypes
Optotypes available in this option are much bigger than standard ones, of visus
units up to ft 20/1600
1.
Enter ‘
AMD Optotypes
’, using Arrow Up or Arrow Down button.
2.
Press Right Arrow to enter edit mode.
3.
Use Up and Down Arrows to set edited parameter to ‘
Yes
’ or ‘
No
’ value.
4.
Press Left Arrow to apply changes.
HINT !