DIGITAL WEIGHING INDICATOR TI-5700
- 23 -
Accumulation & hold operation
◊ F16: Accumulation mode (0~3)
Value
Description
0
Manual (+ only)
The selection for the way of adding data.
1
Manual (+/-)
2
Automatic (+ only)
3
Automatic (+/-)
◊ F17: Set inhibition range for accumulation (0~4)
Value
Description
0
Add all of the data
In automatic mode, data can be accu-
mulated after the weight data enters
the inhibition range.
Automatic accumulation does not work
in the 0.
1
less
than or equal to 5d
2
less
than or equal to 10d
3
less
than or equal to 20d
4
less
than or equal to 50d
◊ F18: Hold mode (0,1)
Value
Description
0
Average hold
Compute the average value.
1
Peak hold
Measure maximum value and hold on display
◊
F19: Average hold time (0~9)
Value
Description
0
Sampling hold
Averaging time by one step.
(1: Very fast ~ 4: Normal ~ 9: Very slow)
4
Normal speed
9
Slow speed
◊
F20: Initialization average hold weight (0~9)
Value
Description
0
Zero
4
Below 4 division
9
Below 9 division