DIGITAL WEIGHING INDICATOR    TI-5700 

 

- 18 - 

6. Setting Mode 

 

 

6.1. How to enter 

To enter the function settings, do either of the following. 
A) When the display is off, press ON/OFF key while pressing the TARE key. 
B) When in the weighing mode, press and hold the TARE key for more 2 seconds.

 

 

Step 1. At this time, 

F-01

” is shown on the display after “

SET

” message. 

Step 2. You can select the menu that you want to set. Enter number of set menu 

by pressing the arrow keys and then press ENTER key. 

 

6.2. Available keys 

  : Used to move the input value to the left or right by one place. 

  : Used to increase the setting constant one by one. 

  : Used to escape from setting mode. 

  : Used to move next menu after completing input value. 

 

6.3. Function menu 

 

  General function                                                                                                                 

 

 
 

 

 

F01: A/D output data rate (0~9)   

Value 

Description 

00 

  256Hz 

If a higher parameter is set, the display speed will be   

slow, but response will be more insensitive to   
environment such as vibration. 

In contrast, If a lower parameter is set, the display   
speed will be fast, but response will be more sensitive to   
environment such as vibration. 

01 

  128Hz 

02 

  64Hz 

03 

32Hz 

04 

16Hz 

05 

10Hz 

06 

8Hz 

07 

6Hz 

08 

4Hz 

09 

2Hz 

 
 

Bolded display is factory default set value. 

 

Summary of Contents for TI-5700

Page 1: ...TI 5700 VER 101 WEIGHING INDICATOR OPERATION MANUAL...

Page 2: ...g or missing parts of the contents listed in this manual please contact us 4 To improve the product performance functions can be changed with no notice 5 Please understand that TMT does not have respo...

Page 3: ...check batching mode 36 9 Test Mode 44 10 Solid State Relay and Control Inputs OP 01 46 11 RS 232C Interface OP 02 47 12 RS 422 485 Interface OP 03 51 13 Analog Output OP 04 52 14 Current Loop Output O...

Page 4: ...tion should be conducted under our surveillance Do not install the indicator in strong direct sunlight and dust Please confirm that the local voltage is correct for the indicator Do not use inflammabl...

Page 5: ...w OP 01 3 solid state relay and inputs OP 02 RS 232C serial interface OP 03 RS422 485 serial interface OP 04 4 20mA 0 10V analog output OP 05 Current loop output OP 06 LAN IEEE 802 3 OP 07 BCD output...

Page 6: ...data 10 Unit Unit used to weigh 11 Stable Indicates when the weight is stable 12 USB USB host port for memory stick 13 CAL cover CAL switch cover For calibration 14 ON OFF key Power ON OFF of indicato...

Page 7: ...nt cover 2 Panel bracket Standard panel bracket 3 Option panel cover For OP 01 02 03 04 05 06 and 07 4 Load cell connector Connect the load cell 5 Ground terminal Frame ground F G 6 DC jack Uses the s...

Page 8: ...may invalidate any approval or warranty and may be dangerous For availability of approved enhancements please check with your dealer When the standard AC adapter is used Check the power supply voltage...

Page 9: ...l wire you must power off and be sure to connect the load cell wire to the terminal correctly The output voltage of a load cell is a very sensitive signal Don t weld near the load cell indicator Also...

Page 10: ...r and load cells EV Excitation voltage mV RO Rated output mV V e Weighing interval kg LC Rated capacity of load cell kg N Number of load cells EV RO e 0 2 LC N Input sensitivity is 0 2uV D EXAMPLE Loa...

Page 11: ...3 Align the option board with guides inside of option part and Insert the option board to the connector of main board Step 4 Tighten the screws of option board 4 4 Mount bracket install OP 12 Step 1...

Page 12: ...ress ENTER key to save and move to next step If you press the ESC key it turns back to weighing mode after save maximum capacity minimum scale division The display resolution is cannot be over than 50...

Page 13: ...nce weight is shown on the display Confirm if the displayed weight is equal to the balance weight that you have set in step 5 The bias is 0 when the OK lamp lights up If HI or LO lamp is on you can be...

Page 14: ...ey it turns back to weighing mode after save maximum capacity minimum scale division The display resolution is cannot be over than 50 000 Display resolution Max capacity Min scale division 4 Setting c...

Page 15: ...increase the digit increase the value Press ENTER key to save and move to next step 3 Setting minimum scale division After DIV is displayed select min scale division with keys increase the value decr...

Page 16: ...s turned on Do not press the keys while the stable mark is off 7 2nd 3rd 4th 5th input balance weight span calibration Repeat steps 5 6 7 for each 5th set value must be the maximum capacity If you wan...

Page 17: ...ed 2 Setting gravity acceleration of calibration location After GSET 1 is displayed input gravity acceleration of calibration location with keys increase the digit increase the value Press ENTER key t...

Page 18: ...ble keys Used to move the input value to the left or right by one place Used to increase the setting constant one by one Used to escape from setting mode Used to move next menu after completing input...

Page 19: ...2 5d 2sec F03 Motion detection 0 10 Value Description 00 OFF Per each A D conversion the current weight is compared to previous weight by this parameter If any of the set parameters are exceeded stabl...

Page 20: ...ways ON F07 Backlight brightness level control 0 9 Value Description 0 10 The brightness of the backlight is adjusted depending on this parameter 4 50 9 100 F08 Weighing unit change 0 2 Value Descript...

Page 21: ...ss output 3 Print data output Included in ZigBee wireless output 4 Accumulation 5 Setting item code 6 Start batching 7 Stop batching 8 Hold start release 9 Data output to USB memory stick F12 F2 key f...

Page 22: ...ireless output 7 Print data output Included in ZigBee wireless output 8 Start batching 9 Stop batching 10 Net weight display while shorting the terminal 11 Accumulated data display while shorting the...

Page 23: ...he inhibition range Automatic accumulation does not work in the 0 1 less than or equal to 5d 2 less than or equal to 10d 3 less than or equal to 20d 4 less than or equal to 50d F18 Hold mode 0 1 Value...

Page 24: ...tching 2 6 Loss in weight batching 7 Built in automatic program Normal batching 1 8 Normal batching 2 9 Loss in weight batching F22 Automatic tare when start for F21 7 8 9 0 1 Value Description 0 Manu...

Page 25: ...Item code number 0 7 Value Description 0 Item code number 0 3 Item code number 3 7 Item code number 7 F27 High limit weight for F21 1 9 F28 Low limit weight for F21 1 9 F29 Zero band for F21 1 9 F30 F...

Page 26: ...1 bit Non parity 1 Data 7 bits Stop 1bit Even parity 2 Data 7 bits Stop 1bit Odd parity F35 Data format 0 1 Value Description 0 Format 1 Transmit 18 bytes Refer to 11 RS 232C Interface 1 Format 2 Tran...

Page 27: ...F39 Accumulated data serial output 0 2 Value Description 0 No output 1 Manual mode 2 Automatic mode F40 Printing format 0 2 Value Description 0 Printing format 1 1 Printing format 2 2 Printing format...

Page 28: ...en parity 2 Data 7 bits Stop 1bit Odd parity F44 Data format 0 1 Value Description 0 Format 1 Transmit 18 bytes Refer to 11 RS 232C Interface 1 Format 2 Transmit 22 bytes F45 Command address ID 00 99...

Page 29: ...232C Interface 5 Automatic print mode Transmit only in stable state F48 Accumulated data serial output 0 2 Refer to 7 3 Accumulation Value Description 0 No output 1 Manual mode 2 Automatic mode F49 Pr...

Page 30: ...de 0 1 Value Description 0 0 24mA current mode 1 11 11V voltage mode F55 Weight value at 4mA or 0V output 0 to 999999 Value Description 0 Input the weight value at 4mA or 0V 999999 The decimal point i...

Page 31: ...isplayed weight 1 Gross weight 2 Net weight F59 Output mode 0 1 Value Description 0 Not used 1 To USB memory stick Refer to F11 F12 of setting mode Wireless communication F60 Output mode 0 3 Value Des...

Page 32: ...net value with the container weight subtracted from the total weight if you place an object into a container to weigh it Weighing Tare Step1 Place the tare on the weighing tray Step2 Press the TARE k...

Page 33: ...count and total The limitation of accumulation count and total are 999999 Action of accumulating data When accumulating data the display blinks once If the accumulated data is stored the mark turns on...

Page 34: ...d data and count are cleared Output of accumulated data The accumulated data can be output to the serial interface Refer to F38 F47 Step1 Press and hold the ACCUMULATION key for more 2 seconds TOTAL i...

Page 35: ...ay shows automatically and average weight will appear If the weight fluctuation range exceeds within 10 of weight the scale will retry auto hold average Releasing the data when a displayed value small...

Page 36: ...and is not lost even if power is off Setting the parameters Refer to F25 F31 of setting mode Code memory can be selected by F25 of setting mode external control input or a command via the serial inte...

Page 37: ...er mode F21 2 When the scale is stable LO HI and OK relays is ON and it is OFF after 1 5sec The operation of the display lamp and the relay are the same Limit type checker mode F21 3 The operation of...

Page 38: ...Over It is ON when the net weight is greater than the weight Final High limit after Final output is ON and HI lamp turns on 4 Under It is OFF when the net weight is greater than the weight Final Low l...

Page 39: ...ay shows 0kg 2 Preliminary It is ON when the net weight is greater than the weight Final Preliminary 3 Final It is ON when the net weight is greater than the weight Final Free fall 4 Zero band It is O...

Page 40: ...r than Full weight 2 Press TARE key so that the display shows 0kg 3 Final It is ON when the net weight is greater than the weight Final Free fall 4 Over It is ON when the net weight is greater than th...

Page 41: ...l High limit after Final output is OFF and HI lamp turns on 5 Under It is OFF when the net weight is greater than the weight Final Low limit after Final output is OFF and LO lamp turns on 6 Finish Whe...

Page 42: ...greater than the weight Final Preliminary 4 Final It is OFF when the net weight is greater than the weight Final Free fall 5 Zero band It is ON when gross weight is lower than zero band value 6 Stop i...

Page 43: ...the net weight is greater than the weight Final Free fall 5 Over It is ON when the net weight is greater than the weight Final High limit after Final output is OFF and HI lamp turns on 6 Under It is O...

Page 44: ...o escape from test mode Used to move next menu Perform the test 9 3 Test menu TEST 1 Keyboard test Display Description NO 1 If you press the key that you want to test the key number is appeared on the...

Page 45: ...on 00 00 31 01 Wait for transmission and reception Transmit 0x01 ZERO key Receive 0x31 No 1 key in PC Do this test in hyper terminal of PC after the connecting COM port of PC and RS232C port of indica...

Page 46: ...nal control functions at F13 15 of the setting mode When connecting each function pin to the common pin the indicator makes the action Please make the minimum time to input a data with over 50msec 10...

Page 47: ...5P 6P 7P 8P 9P N C RXD TXD N C GND N C 11 2 Data format There are two type of data format set at setting mode F34 or F43 Format 1 18 bytes A D company standard Header 1 Header 2 Data Unit Terminator S...

Page 48: ...0x55 0x53 OL Overload 0x4F 0x4C HD Hold weight data 0x48 0x44 Header 2 GS Gross data 0x47 0x53 NT Net data 0x4E 0x54 ID Command address ID is selected in F35 F44 If the ID is 13 0x13 is sent Data Exam...

Page 49: ...de F35 44 The reply Data or error code is also sent with the address If the ID is 13 command is 1 3 R W When the received command can t be performed such as the busy state the code I is sent back STX...

Page 50: ...CA CACRLF Changing the Code Memory 0 7 e g code memory 3 C3 C3CRLF Commands to set points Function Command Response High limit weight e g item code 3 25 5kg S3 1 000255 S3 1 000255CRLF Low limit weig...

Page 51: ...nous transmission format Baud rate 2400 4800 9600 19200 38400 57600 115200 bps Data bit 7 8 bits Parity bit Even or Odd For 7 data bits or Non parity For 8 data bits Start bit 1bit Stop bit 1bit Code...

Page 52: ...mA 0 10V Resolution 16bits Full scale error 0 001 FSR Temperature coefficient 1ppm FSR C 13 2 Pin connection Adaptable connector D sub 9 pin connector Female Pin Name Description 1 VS 0 10V Sense 2 VO...

Page 53: ...as the RS232C data format Refer to 11 RS232C Interface However Baud rate is 2400bps Adaptable connector D sub 9 pin connector Female 1P 2P 3P 4P 5P 6P 7P 80 9P N C CL1 N C CL2 N C 15 LAN IEEE 802 3 O...

Page 54: ...ansmits 0000 0000 0011 0111 1000 0010 Inner circuit is electronically disconnected by photo coupler The output circuit is an open collector type If you hook up this line to TTL please use external VCC...

Page 55: ...C 8 3x10 33 N C 9 4x10 34 N C 10 1x100 35 N C 11 2x100 36 N C 12 3x100 37 External VCC 13 4x100 38 N C 14 1x1 000 39 External VCC 15 2x1 000 40 N C 16 3x1 000 41 N C 17 4x1 000 42 High Low 18 1x10 000...

Page 56: ...ormat Baud rate 9600 bps Data bit 8 data bit Non parity bit Start bit 1bit Stop bit 1bit Code ASCII USB 2 0 compatible FAT32 file system Aaptable connector USB A type Female 17 2 Data format There are...

Page 57: ...el file format Format 2 22 bytes F57 2 22 bytes CAS company standard Refer to 10 2 Data format 17 3 Command mode Same as RS232C data format Refer to 10 3 Command mode Header ST Stable weight data US U...

Page 58: ...f other brand s sub display must be used TF200 ZigBee dongle PC or PLC Must be used TF200 ZigBee dongle Specifications RF frequency range 2400 2483 5 MHz Output power Max 4dBm Channel width 2 MHz Freq...

Page 59: ...to additional manual Specifications Bluetooth specification Fully qualified with Bluetooth v2 0 EDR specification EDR Enhanced data rate compliant with v2 0 of specification for both 2Mbps and 3Mbps R...

Page 60: ...Step 3 If adaptor is still being charged the RED lamp is on Step 4 If a charging is completed the RED lamp is off Step 5 The battery charging time takes about 6 hr Charging time is subject to be chang...

Page 61: ...range 0 39mV Non linearity 0 0015 of FSR max Load cell excitation DC 5V 5 300mA Up to 8 unit of 350ohm Conversion method Sigma delta External resolution Up to 100 000counts Output data rate 4 7 to 200...

Page 62: ...DIGITAL WEIGHING INDICATOR TI 5700 62 Dimensions...

Page 63: ...clude external key for long time or having problems in key part Make sure there are no problems with key switch part OVER The gross weight is over maximum capacity Don t load the article whose weight...

Page 64: ...DIGITAL WEIGHING INDICATOR TI 5700 64 MEMO...

Reviews: