Connecting cable (option "mass pressure-/temperature-compensated")
Order code for "Sensor version; DSC sensor; measuring tube", option
Standard cable
[(3 × 2) + 1] × 0.34 mm
2
(22 AWG)PVC cable with common shield
(3 pairs, pair-stranded)
1)
Flame resistance
According to DIN EN 60332-1-2
Oil-resistance
According to DIN EN 60811-2-1
Shielding
Galvanized copper-braid, opt. density approx. 85%
Cable length
30 m (98 ft)
Operating temperature
When mounted in a fixed position: ‒50 to +105 °C (‒58 to +221 °F);
when cable can move freely: ‒25 to +105 °C (‒13 to +221 °F)
1)
UV radiation may cause damage to the outer jacket of the cable. Protect the cable from exposure to sun
as much as possible.
Overvoltage protection
The device can be ordered with integrated overvoltage protection for diverse approvals:
Order code for "Accessory mounted", option NA "Overvoltage protection"
Input voltage range
Values correspond to supply voltage specifications → See 20
1)
Resistance per channel
2
·
0.5 Ω max.
DC sparkover voltage
400 to 700 V
Trip surge voltage
< 800 V
Capacitance at 1 MHz
< 1.5 pF
Nominal discharge current
(8/20 μs)
10 kA
Temperature range
‒40 to +85
°
C (‒40 to +185
°
F)
1)
The voltage is reduced by the amount of the internal resistance I
min
·
R
i
Depending on the temperature class, restrictions apply to the ambient temperature for
device versions with overvoltage protection .
Performance characteristics
Reference operating
conditions
• Error limits following ISO/DIN 11631
•
+
20 to
+
30
°
C (
+
68 to
+
86
°
F)
• 2 to 4 bar (29 to 58 psi)
• Calibration system traceable to national standards
• Calibration with the process connection corresponding to the particular standard